83
3. Save the result
Same as Section 5.1.2 Pressure transmitter.
5.2 Task
The calibrator provides task functions which automate preset calibration routines, and collect, record, and save test
data.
Enter the task interface: Click
, and select the task in the menu .
Task list interface: Click
on the right side of the screen to create a new task, click
to delete tasks, and
click
to create a new folder.
Test parameter interface: Click
on the right side of the screen to delete the task, click
to edit the task,
click
to copy the task, and click to start the task.
5.2.1 Analog pressure gauge
1. Connection
Connect the calibrator and the analog pressure gauge as shown in Figure 3-14.
It is recommended to connect the reference port (REF) when calibrating a differential pressure device.
2. New task
Select the “analog pressure gauge” icon
in the test type list, and input the information of the pressure gauge.
Содержание 761A
Страница 1: ...761A Automated Pressure Calibrator...
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 16: ...X Table 5 13 Task mercury barometer 120 Table 5 14 Task barometric altimeter 123...
Страница 19: ......
Страница 20: ......
Страница 30: ...10 Warranty 1 year Optional Software ACal and Additel Land Additel Link for access via mobile application...
Страница 31: ...11 2 Installation 2 1 Features 2 1 1 Basic structure Figure 2 1 Basic structure...
Страница 33: ...13 LAN Interface USB port Power supply input Label Battery...
Страница 47: ...27 Figure 3 3 Gauge pressure outlet...
Страница 137: ...117 Figure 5 9 Calibration of the aneroid barometer Figure 5 10 Calibration of the barometric altimeter...
Страница 151: ...131...