9
1.2.2 Signal measure specification (environmental temp.: 20±10
℃
, 1 year accuracy
Table 2 Signal measure specification
Specification
Range
Accuracy
Resolution
Note
Voltage Measurement
±300mV
0.015%RDG + 0.005%FS
1uV
Impedance:
>100MΩ
±30V
0.015%RDG + 0.005%FS
0.1mV
Impedance:
>1MΩ
Current Measurement
±30mA
0.015%RDG + 0.005%FS
0.1uA
Impedance: <
40Ω
Frequency
Measurement
0.01~50000Hz (auto range)
0.005% RDG + 2 last digit
6-digit auto-resolution
Min threshold voltage:
2.5V
Units: Hz, kHz, MHz, CPM, CPH, s, ms, us
Switch On-Off
Measurement
Inspection voltage: (3 ~ 30)V
Response speed: < 10ms, supports wet and dry switch
Pulse Count
0 ~ 9999999, optional rising edge and
falling edge Min threshold voltage: 2.5V
Loop Power
20V ± 10%, max output impedance: 320Ω, max load current:
25mA
Note 1
:
When the environmental temperature is (-20~-10) °C, the temperature coefficient is:
(1) Measure of voltage, current
:
±5ppmFS/°C.
Note 2
:
Input characteristics:
(1) Voltage range
:
-
300~300mV input impedance >100 MΩ;
(2) Voltage range
:
-
30~30V input impedance >1MΩ;
(3) Current measure
:
input impedance <40Ω;
Note 3
:
Minimum threshold voltage for frequency and pulse measure
:
2.5V;
Note 4
:
The frequency measure supports following units: Hz, kHz, MHz, CPM, CPH,s, ms,
μ
s;
Note 5
:
There are rising edge and falling edge trigger mode available for pulse measure.
Содержание 273E Series
Страница 1: ...1 273Ex Handheld Pressure Calibrator...
Страница 2: ......
Страница 3: ......
Страница 4: ...ADT273Ex Handheld Pressure Calibrator User Manual Version 2302V02 Additel Corporation...
Страница 6: ......
Страница 20: ...10 1 3 Basic Structure Figure 1 Basic Structure...
Страница 41: ...31 Figure 16 Sensor trim Figure Figure 17 Process variable selection...
Страница 56: ...46...