41
Zero offset
Zero offset generated by performing
zeroing
Zero offset when performing differential pressure module
zeroing
Pressure module A
S/N of the pressure module
Pressure module B
S/N of the pressure module
Zero
Perform zero to the differential pressure module, the zero
offset will be saved and used for calculation
Cancel zero
Cancel the zero-offset created when performing the zero, reset
the zero offset to 0
7.2 Units converter
The device provides unit converter for pressure units and temperature units, users can select multiple pressure units or
temperature units for conversion. Click the unit converter icon on the main interface of the device to enter the unit
converter. Click
icon on the right top to select and switch between different unit types. Click the unit under the value
display to select and switch between different units.
7.3 Wiring help
The wiring help function in the device can help users to connect the wires correctly under various working conditions of
the calibrator. It can be entered by clicking the "Wiring help" menu in each channel, or by clicking the "wiring help" icon in
the main interface of the device.
7.4 Leak test
The device provides a pressure leak test function, which can perform self-checking on the seal of the device. Click on