3.3.5 Zeroing
See paragraph 3.1.8.
3.3.6 Statistics
See paragraph 3.1.10.
3.4 HART calibration
The device provides HART bus communication, using the simplified DD file, provide general settings, service and
calibration for HART temperature and pressure transmitters. It is recommended to read the transmitter’s manual before
use.
Note: the calibrator will always be the master during communicating with the HART device. In order to avoid harm to
the controlling system, please separate the HART device from the controlling system before connecting the calibrator with
the HART device.
3.4.1 Search and connection
Switch to HART on the Calibrator page, the calibrator will automatically use the last selected power supply
configuration (default is the internal resistance and internal source), and search for the address "0". After searching for the
HART device, it will automatically connect and display its value. Click the menu icon
to select search, or click the
Figure 12 External external resistor
(supports loop current measurement)
Figure 11 External external resistor
Figure 10 Internal internal resistor