3.1.7 Scaling
The scaling function gives the calibrator the ability to convert electrical measurement signals into other signals. Click
the menu button
of the measurement channel, select the scaling menu in the pop-up menu, and the parameter
configurations related to scaling will be displayed (see Table 8).
Table 8 Scaling configurations
Subject
Valid Value
Description
Enable/disable
Enable/disable
Enable or disable the scaling function
Input range
(
-9999999~9999999
)
Range before scaling
Output range
(
-9999999~9999999
)
Range after scaling
Resolution
0/0.1/0.01/0.001
The number of decimal places displayed after scaling
Transfer function
Linear, square or square root
Scaling transfer function
3.1.8 Filter
The calibrator provides two filtering methods: first-order linear filtering and moving average filtering to process data to
meet the needs of different usage scenarios. Click the menu button
of the measurement channel, select the filter
menu in the pop-up menu, and the filter setting interface will be displayed (see Table 9).
Table 9 Filter configurations
Subject
Valid Value
Description
Enable/disable
Enable/disable
Enable or disable the filter function
Method
first-order linear filter or moving average filter
Filtering algorithm
Filter coefficient
0.01 ~ 1
Only available when the first-order linear filter is selected
Содержание 226
Страница 1: ...226 227 Multifunction Process Calibrator...
Страница 2: ......
Страница 3: ...ADT226 227 Multifunction Process Calibrator User Manual Version 2306V01 Additel Corporation...
Страница 4: ......
Страница 6: ......
Страница 31: ...19 1 4 Basic Structure Figure 1 Basic Structure...
Страница 33: ...21 Figure 2 Power adaptor...
Страница 64: ...52 External module information including external module A and external module B...
Страница 107: ...95...