8250A Optical Power Meter Operation Manual
6.2 Performance Test of the Optical Sensors
6-5
6.2
Performance Test of the Optical Sensors
To calibrate the optical sensors, the light source, optical fiber, and standard optical power meter, all of which
satisfy the specifications described below, as well as a measuring environment with little or no vibration, at
a temperature of +23
C
5
C, and a temperature change of
1
C are required.
To evaluate the optical sensor with the calibration wavelength addition option, prepare the relevant light
source and fibers.
Table 6-4 Equipment Required for Evaluating Optical Sensors
Product
Light source
Fiber
Sensor adapter
Wavelength
Output
Stability
82311B
780 nm
1 mW
0.20%
PMF/SMF or GI50um
WBL-Q17ADAPTER
82312B
405 nm
1 mW
0.20%
PMF/SMF
WBL-Q82312ADAPTER
82313B
650 nm
1 mW
0.20%
PMF/SMF
WBL-Q82313ADAPTER
82321B
780 nm
1 mW
0.20%
PMF/SMF or GI50um
A08012
82322B
405 nm
1 mW
0.20%
PMF/SMF
A08012
82323B
650 nm
1 mW
0.20%
PMF/SMF
A08012
82314B
405 nm
1 mW
0.20%
PMF/SMF
WBL-Q82312ADAPTER
82314BW
405 nm
1 mW
0.20%
PMF/SMF
WBL-82314WADAPTER
82324B
405 nm
1 mW
0.20%
PMF/SMF
A08012
Optical power meter 8230/8250A
Must be calibrated at an accuracy of 0.8% or less for
each wavelength.