8230E Optical Power Meter Operation Manual
6.2 Measuring Instruments and Other Equipment Required to Check the Performance of the Optical Sensor
6-5
6.2
Measuring Instruments and Other Equipment Required to Check the Performance
of the Optical Sensor
To calibrate the optical sensors, the light source, optical fiber, standard optical power meter, all of which meet
the specifications described below, as well as a measuring environment with little or no vibration, at a tem-
perature of +23
C
5
C, and a temperature change of
1
C are required.
Table 6-4 Equipment Required for Evaluating Optical Sensors
Product
Light source
Fiber
FC adaptor
Wavelength
Output
Stability
82311B
780 nm
1 mW
0.20%
PMF/SMF or GI50
m WBL-Q17ADAPTER
82313B
650 nm
1 mW
0.20%
PMF/SMF
WBL-Q82313ADAPTER
82321B
780 nm
1 mW
0.20%
PMF/SMF or GI50
m A08012
82323B
650 nm
1 mW
0.20%
PMF/SMF
A08012
8230E/8230/8250A
For connecting a sensor
Optical power meter Standard
Must be calibrated at an accuracy of 0.8% or less
for each wavelength.