8230E Optical Power Meter Operation Manual
3.5 Wavelength Preset
3-6
3.5
Wavelength Preset
Frequently used wavelengths can be set easily by registering the four wavelengths. The preset wavelength
cannot be set and called by using the remote command.
3.5.1
Setting the Preset Wavelength
The four wavelengths from No.0 to No.4 can be set.
The wavelength setting range is from 0 to 9999.
An example, in which No.1 is set to 405 nm, is shown below. See Figure 3-2.
Key operation
Display
1.
[MENU]
[S.CNT]
[010]
The first item of the menu is displayed.
2.
[PREV]
3 times
[P.SET]
The current setting of the No.0 preset wavelength
is displayed.
[0 0780]
The least significant digit blinks. (Cursor)
3.
[HOLD]
[P.SET]
[0 0780]
The most significant digit blinks. (Cursor moves)
4.
[ ]
[P.SET]
[
]
[1 0780]
The current setting of No.1 is displayed.
5. [HOLD]
[P.SET]
[1 0780]
The least significant digit blinks. (Cursor moves)
6.
[ ]
[P.SET]
[
]
[
]
[1 0405]
Move the digit and increase or decrease the value
[
]
to enter 405.
7.
[ENTER]
Normal display
Setting complete.
Press
[CANCEL]
to cancel the setting.