7451A/7461A/7461P Digital Multimeter Operation Manual
7.1.10 Measurement Time and Display Digits
7-17
7.1.10
Measurement Time and Display Digits
Maximum measurement speed and its conditions
Sampling rate
Measurement speed (In Auto-range OFF)
Integration
time
Auto-zero
DCV
Bch-DCV
DCI
2W
LP-2W
4W
LP-4W
ACV
ACI
Continuity
Diode
ACV(AC+DC)
ACI(AC+DC)
FAST1
1000 times/s *1
110 times/s
400 times/s
4 times/s
0.02 PLC
OFF
FAST2
300 times/s *1
80 times/s
250 times/s
4 times/s
0.1 PLC
OFF
MED
20 times/s
20 times/s
20 times/s
3 times/s
1 PLC
ON
SLOW1
4.5 times/s
4.5 times/s
4.5 times/s
2 times/s
5 PLC
ON
SLOW2
2 times/s
2 times/s
2 times/s
1 time/s
10 PLC
ON
*1
Under the following conditions
Trigger source: BUS, EXT, MAN
Sampling count: 2 or more
Complete signal output mode: SINGLE
Display: OFF, Calculation: OFF, Measurement data memory: ON
7461A/61P
7451A
Measurement speed
20,000 times/s
5,000 times/s
Target function
DCV, DCV-Bch, DCI, 2W
, LP-2W
Integration time
10
s
100
s
Sampling interval
50
s
200
s
Other
*1
*1
Under the following conditions
Auto-range:
OFF
Trigger source:
BUS, EXT, MAN
Sampling count:
2 to 16,000
Complete signal output mode:
SINGLE
Display: OFF
Measurement data memory:
ON
7451A/61A: Up to 10,000 data
7461P:
Up to 16,000 data