7351 Series Digital Multimeter Operation Manual
5.12 System Settings
5-35
5.12 System Settings
5.12.1
Buzzer Setting
Select ON or OFF in ‘BEEP’ in
in
.
ON
The buzzer sounds.
OFF
The buzzer does not sound.
The buzzer does not sound regardless of the result of comparator calculation.
Press
to apply the setting.
If an error occurs, the buzzer sounds even when BEEP is set to OFF.
5.12.2
Power Supply Frequency
When the sampling rate is set to MED, the effect of induction noise can be reduced by synchronizing the
integration time with the power supply frequency.
Select a power supply frequency according to the area where this instrument is used.
Select the following from 'LINE' in the
category in
.
50
The commercial power supply frequency is 50 Hz.
60
The commercial power supply frequency is 60 Hz.
5.12.3
Disabling Functions
To prevent incorrect functions from being selected because of malfunctions and to use this instrument
safely, some functions can be disabled.
Select each function to be set to ENA or DIS in ‘F.Inhi’ in
in
.
Press
to enter in the entry layer and select ENA (Enabling) or DIS (Disabling) by
using
and .
Press
to apply the setting.
•
The relationship according to disabling the function setting
Although the ENA or DIS can be set for each function, some functions are automatically disabled
depending on the settings of other functions.
If a higher function setting in the dependency relation is disabled, no lower function setting can be
changed.