7351 Series Digital Multimeter Operation Manual
4.2.1 Measurement Functions
4-14
Figure 4-10 Continuity Measurement
4.2.1.7
Diode Measurement
1. Connect a DUT to the V and COM terminals on the front panel.
The diode indicator (
) is activated.
2. Press
and
.
A current of approximately 1 mA is applied from the COM terminal to the V terminal, and the voltage
between both terminals is measured.
Measurement range
2000 mV
Measurement current
1 mA
WARNING:
Do not apply voltage that exceeds the maximum allowable applied voltage. Fire or electric
shocks due to the failure of this instrument may occur.
Table 4-7 Maximum Allowable Applied Voltage (Continuity)
Between
and COM terminals
1000 Vpeak
Between COM terminal and chassis (grounding)
500 V
Table 4-8 Maximum Allowable Applied Voltage (Diode)
Between V and COM terminals
1000 Vpeak
Between COM terminal and chassis (grounding)
500 V
Current
SHIFT
DCI