7351 Series Digital Multimeter Operation Manual
4.2.1 Measurement Functions
4-13
Figure 4-9 AC Current Measurement (ACI and ACI (AC+DC))
Measurement method
TrueRMS
Input range
5% or more of a full scale
Crest factor
3 (This is restricted to the current protection fuse rating.)
Filter response time
Approx. 1 sec (Time until the measurement value reaches within 0.1%
of the final value)
Auto-range restrictions
The 10 A range is a fixed range. The auto-range cannot function between the 10 A range and 200 mA,
2000 mA ranges.
4.2.1.6
Continuity Measurement
1. Connect a DUT to the
and COM terminals on the front panel.
The continuity indicator (
) is activated.
2. Press
and .
The resistance is measured in 2W
. If the measured resistance is lower than the continuity threshold
constant, the buzzer sounds.
Measurement range
2000
Measurement current
1 mA
Continuity threshold constant 1
to 1000
WARNING:
Do not apply voltage that exceeds the maximum allowable applied voltage. Fire or electric
shocks due to the failure of this instrument may occur.
10 A or less
10 Arms or less
250 V or less
2 A or less
2 Arms or less
SHIFT
FREQ