6247G/6247C DC Voltage Current Source/Monitor Operation Manual
3.1 Diode Measurement
3-2
Connecting the DUT
1. Connect the A08532 alligator clips to the supplied A01044 input and output
cables and connect the diode as shown below.
A 2-wire connection is used in this measurement example.
Figure 3-1 Diode Measurement Connection
Measuring the diode forward voltage
2. Set the conditions for VF measurement.
3. Short-circuit the HI OUTPUT cable (red) and the LO OUTPUT cable (black) to
obtain the NULL value.
4. Set the output status to Operate.
5. Press
NULL
.
6. Set the output status to Standby or Suspend.
7. Connect the HI OUTPUT cable (red) to the anode of the diode and the LO OUT-
PUT cable (black) to the cathode.
8. Set the output status to Operate.
The VF measurement result is displayed.
Diode
Red
Black