6247G/6247C DC Voltage Current Source/Monitor Operation Manual
2.1.2 Screen Display (Annotations)
2-7
2.1.2
Screen Display (Annotations)
This section describes the screen display (annotations).
Figure 2-11 Screen Display (Annotations)
1. Source value
Displays a voltage source (VS) or current source (IS) value with
a unit.
2. Measurement value
In Operate:
Displays a measurement value.
In Standby or Suspend:
Displays the suspend voltage. Displays the output impedance
in Suspend status.
HZ: High impedance status
LZ: Low impedance status
(For more information on this operation, refer to Section 5.2,
3. Measurement function
Displays the measurement function.
I:
Current measurement (IM)
V:
Voltage measurement (VM)
R:
Resistance measurement (RM)
_ :
Measurement OFF
4. Left status indicators
PLS:
Pulse source mode.
SWP:
Sweep source mode.
NOTE:
PLS+SWP means pulse sweep, and DC+SW means DC
sweep.
DC:
DC source mode.
AC:
(Not in use)
HOLD: The trigger mode is set to HOLD.
2
1
3
4
8
11
9
5
7
6
10