6247G/6247C DC Voltage Current Source/Monitor Operation Manual
8.5.3 Voltage Measurement Calibration
8-14
8.5.3
Voltage Measurement Calibration
1. Move to the voltage measurement calibration mode by XVM.
2. Select the range.
5 V range:
XR4
50 V range:
XR5
250 V range:
XR6
3. Enter the DMM data input mode by XDAT.
4. Set the DMM read value by XD data.
5. Read out and check the measurement data.
6. Move to the full-scale calibration mode by XNXT.
7. Set the DMM read value by XD data.
8. Read out and check the measurement data.
9. Move to the next step.
When changing the voltage range:
XNXT
When moving to the current source or voltage limiter calibration:
XIS, XVLH, XVLL
8.5.4
Current Source/Current Limiter Calibration (3
A Range)
1. Select the current calibration mode.
Current source:
XIS
Current HI limiter:
XILH
Current LO limiter:
XILL
2. Select the 3
A range.
3
A rang:
XR-2
3. From this step, perform the same procedure as described in Section 8.5.2, "Volt-
age Source/Voltage Limiter Calibration."
4. Move to the next step.
When changing the current calibration mode:
XNXT
When moving to the current measurement calibration: XIM