6247G/6247C DC Voltage Current Source/Monitor Operation Manual
7. PERFORMANCE TEST
7-1
7.
PERFORMANCE TEST
This chapter describes the methods for checking whether the 6247G/6247C operates in the specified accu-
racy.
7.1
Measuring Instrument Necessary for Performance Test
Use a measuring instrument for the performance test which meets the specifications shown in Section 8.1,
"Measuring Instrument and Cables Necessary for Calibration."
7.2
Connections
The connections required for the performance test are the same as shown in Figure 8-1, "Connections for Cal-
ibration."
7.3
Test Method
Execute the performance test under the following conditions in a location free of dust, vibration, noise or
other adverse conditions:
Temperature:
23
5
C
Relative humidity: 70% or lower
Warm-up:
60 minutes or longer
Self-test, display, key, and buzzer tests
1. Press the
MENU
key to select
SYSTEM
and refer to Section 5.2.14, "Self Test"
to execute.
NOTE:
If an error occurs during the test, refer to Section 5.2.14, "Self Test" to verify
the content of the error.