6247G/6247C DC Voltage Current Source/Monitor Operation Manual
6.7.4 Remote Command List
6-47
Command compatible with former models
Item
Command
Description
Default
Operation
Power
ON
Default
setting
During DC/
pulse
operation
During
sweep
operation
Source Function and
source range
V4
Voltage source function 5 V range
Sus-
pended
when
executed
V5
Voltage source function 50 V range
V6
Voltage source function 250 V range
I-2
Current source function 3 µA range
I-1
Current source function 30 µA range
I0
Current source function 300 µA range
I1
Current source function 3 mA range
I2
Current source function 30 mA range
I3
Current source function 300 mA range
V?
Response: V4
to V6 or I-2
to I3
I?
Source value
(pulse value)
and limit value
D
data UNIT
The source value setting is different depending on
whether UNIT is specified or not.
With UNIT:
Automatically sets the optical range.
Available unit: mV, V,
A ,mA, A
Without UNIT:
Sets the present function and range.
When a unit that is different from that of the
present source function is set, the input data is set
as limit value as follows.
+data: HI limit value
-data: LO limit value
D?
Response: D
<data1>UNIT,D <data2>UNIT
data1:Voltage or current source value
<d.ddddE
d>
*1
data2: Voltage or current limit value
(The polarity is space.)
<0d.dddE
d>
*1
UNIT: V or A
Note:
When the absolute values of HI and LO limit values
are different, the response becomes
D
d.ddddE
dUNIT, D 09.999E + 9UNIT.