6240B DC Voltage Current Source/Monitor Operation Manual
2.2.8 Random Sweep Source
2-47
2.2.8
Random Sweep Source
This section describes an example of random sweep from 1 V to 3 V as shown below.
Turning off measurement enables high-speed sweep.
Figure 2-18 Random Sweep Source
Source mode:
DC sweep
Sweep type:
Random sweep
Bias value:
0 V (default)
Start address:
0
Stop address:
2
Sweep range:
Fixed (3 V range)
Random memory:
1 V, 2 V, 3 V in order from address 0
Measurement:
OFF
Hold time (Th):
1 ms
Source delay time (Tds):
0.03 ms (default)
Measurement delay time (Td):
0.05 ms
Period time (Tp):
2 ms
Current limit value:
1 A (default)
Press TRIG
Value at address 2
Value at address 0
Bias value