8.5.1 Overall Calibration Procedure
8-11
8.5.1
Overall Calibration Procedure
This section describes the overall calibration procedure according to Figure 8-3 Calibration Procedure (1).
1. Enter the calibration mode by CAL1.
2. When executing all the calibrations, initialize calibration data only once by XINI
at the start of the calibrations.
3. When executing voltage calibration, make connection in reference to Figure 8-1
(a). When executing curernt calibration, make connection in reference to Figure
8-1 (b).
4. Press OPR to set the output status to Operate in the calibration mode.
5. Execute voltage calibration according to Figure 8-4 Calibration Procedure (2)
and Figure 8-5 Calibration Procedure (3), or execute current calibration accord-
ing to Figure 8-6 Calibration Procedure (4) and Figure 8-7 Calibration Procedure
6. Press SBY to set the output status to Standby.
7. Store the calibration data to the non-volatile memory by XWR.
8. Finish the calibration mode by CAL0.
8.5.2
Voltage Source/Voltage Limiter Calibration
1. Select the voltage calibration mode.
Voltage source:
XVS
Voltage HI limiter:
XVLH
Voltage LO limiter:
XVLL
2. Select the range.
300 mV range:
XR3
3 V range:
XR4
15 V range:
XR5
3. Enter the DMM data input mode by XDAT.
4. Set the DMM read value by XD data.
5. Move to the full-scale calibration mode by XNXT.
6. Set the DMM read value by XD data.
7. Finish the DMM data input mode by XNXT.
8. Move to the zero calibration fine adjustment mode by XADJ.
9. Check the zero calibration value.
Fine-adjust the calibration coefficient by XUP and XDN.
10. Move to the full-scale calibration fine adjustment mode by XNXT.