![ADCMT 6240B Скачать руководство пользователя страница 234](http://html1.mh-extra.com/html/adcmt/6240b/6240b_operation-manual_2848265234.webp)
6240B DC Voltage Current Source/Monitor Operation Manual
6.7.4 TER? Command
6-40
6.7.4
TER? Command
The TER? command reads out the self test results.
1. Command response
2. Descriptions of values a, b, c, d and e
The TER register column in Table 5-19 shows the error factors and register values for a, b, c, d and e.
For example, an error in VSVM 3 V +FS of the self test is returned as follows.
00000, 00000, 00016, 00000, 00000
Source Operate/Standby
H
Sets the output to OFF (Standby).
E
Sets the output to ON (Operate).
E?, H?
Returns the present output status.
Response:
Random sweep
memory data
(using D com-
mand)
N [adr]
Random sweep memory data setting starts from
the N command and completes at the P command.
0
*6
P
N<adr>,D<data1><UNIT>,D<data2>
<UNIT>,...,P
adr: Memory address (0 to 7999)
data1:Voltage or current source value
at adr
data2:Voltage or current source value
at adr+1
*6:
Not initialized by the RINI or *RST command.
Item
Command
Description
Default
Operation
Power
ON
Factory
shipment
DC/pulse
OPR/SUS
Sweep
OPR/SUS
Status
E?, H?
Operate
E
Suspend
H
Standby
H
Note:
1. When no source range is specified, the optimal
range is set.
2. Source values different from the present source
function cannot be set.
ddddd,ddddd,ddddd,ddddd,ddddd
a
b
c
d
e