![ADCMT 6240B Скачать руководство пользователя страница 141](http://html1.mh-extra.com/html/adcmt/6240b/6240b_operation-manual_2848265141.webp)
6240B DC Voltage Current Source/Monitor Operation Manual
5.2.4 Operations in Low-Resistance Measurement Pulse Source Mode
5-23
Table 5-9 Operations in Low-Resistance Measurement Pulse Source Mode (1/3)
Operational
condition
Trigger
mode
Description
Operation
Remarks
Operate
ON
AUTO
Executes continu-
ous measurement
with the setting
period time Tp.
Tp: Period time
Tw: Pulse width
Td: Measurement
delay time
Tds:Source delay
time
Tm: Measurement
time
(Integration time
+ Measurement
data processing
time)
Tcn: Operate
processing time
Trc: Range change
processing time
Ts: Polarity
switching time
HOLD
Executes measure-
ment after trigger
input.
Standby
Low-resistance calculation
SYNC OUT
COMPLETE OUT [FRONT]
COMPLETE OUT [END]
COMPLETE OUT [HI/GO/LO]
Operate
Standby
Low-resistance calculation
TRIGGER IN
COMPLETE OUT FRONT]
COMPLETE OUT [END]
COMPLETE OUT [HI/GO/LO]
Operate
Measurement
trigger
SYNC OUT