6240B DC Voltage Current Source/Monitor Operation Manual
4.2.6 MENU Key (Parameter Setting)
4-5
LMT Input
Selects the current limiter HL and LL values setting.
Balance :
The HL and LL values change simultaneously in both
polarities
Separate:
The HL and LL values are separately set.
Response
Selects the source response.
Fast :
Fast mode
Slow:
Slow mode; high stability to a load
SWEEP
Sets the sweep source operation-related parameters.
Sweep Type
Selects the sweep type for the sweep source mode.
Linear:
Linear sweep
Fixed:
Fixed sweep
Random: Random sweep
Linear 2: 2-slope linear sweep
SWP Range
Selects the range function for the sweep source mode.
Auto:
Sweeps in the optimum range at every step from the
start value to the stop value.
Fix:
Sweeps in the lowest range which can output any
source values from the start value to the stop value.
Reverse
Switches reverse mode (round sweep) ON or OFF.
On:
Round sweep from the start value to the stop value and
back to the start value
Off:
One-way sweep from the start value and to the stop
value
Repeat Cnt
Sets the number of times sweep repeats.
When it is set to 0, sweep repeats infinitely.
When it is set between 1 and 1000 and the reverse mode is ON,
each round sweep is counted as 1.
Rtrn Bias
Selects whether sweep stops at the bias value or stop value.
On:
The source value returns to the bias value when sweep
stops.
Off:
The source value stays at the stop value when sweep
stops.
SWEEP VAL
Sets the values used for the sweep source operation.
The setting parameters are different depending on the sweep type.