4601 I-V Meter Operation Manual
8.4.2 Calibration Procedure
8-19
8.4.2.3
Voltage Measurement Calibration
The procedure of voltage measurement calibration is described according to Figure 8-7 Calibration Proce-
dure (3).
1. Move to the voltage measurement calibration mode by XVM.
2. Set the calibration mode to Operate by OPR.
3. Select the range.
5 V range:
XR4
50 V range:
XR5
300 V range:
XR6
4. Enter the DMM data input mode by XDAT.
5. Set the DMM read value by XD data.
6. Read out and check the measurement data.
7. Move to the full-scale calibration mode by XNXT.
8. Set the DMM read value by XD data.
9. Read out and check the measurement data.
10. Move to the next step by XNXT.
11. Set calibration mode to Standby by SBY.
8.4.2.4
Current Measurement Calibration
The procedure of current measurement calibration is described according to Figure 8-8 Calibration Proce-
dure (4).
1. Move to the current measurement calibration mode by XIM.
2. Set the calibration mode to Operate by OPR.
3. Select the range.
300
A range::
XR0
3 mA range:
XR1
30 mA range:
XR2
300 mA range:
XR3
3 A range:
XR4
10 A range:
XR5
4. Enter the DMM data input mode by XDAT.
5. Output the ZERO value from the standard.
6. Set the STD value by XD data.
7. Read out and check the measurement data.
8. Move to the full-scale calibration mode by XNXT.
9. Output the -FS value from the standard.
The full scale is calibrated by the -FS value.
Содержание 4601 I-V Meter
Страница 2: ......
Страница 8: ......
Страница 14: ......
Страница 18: ......
Страница 70: ...4601 I V Meter Operation Manual 3 4 1 Menu Structure 3 18...
Страница 80: ......
Страница 92: ......
Страница 129: ...4601 I V Meter Operation Manual 5 6 1 Synchronous Operation 5 37...
Страница 146: ......
Страница 182: ......
Страница 230: ......