4601 I-V Meter Operation Manual
6.2 GPIB
6-2
6.2
GPIB
6.2.1
Overview
GPIB (General Purpose Interface Bus) allows external control of the 4601 measurement function settings,
measurement parameter settings, and reading measurement data, making it simple to configure an auto-
mated measurement system.
As GPIB signals from the 4601 are electrically isolated inside the unit from the measurement signal sys-
tem, the connection of external devices does not affect the measured values. The remote commands are the
same as with USB.
•
General Specifications
Standard:
IEEE-488.2
Code:
ASCII code
Logic level:
Logical 0 (High)
+2.4 V min.
Logical 1 (Low)
+0.4 V max.
Table 6-1 Interface Function
Code
Function
SH1
Source Handshake capability
AH1
Acceptor handshake capability
T6
Basic Talker, unaddressed if MLA, serial poll
L4
Basic Listener, unaddressed if MTA
SR1
Service Request capability
RL1
Remote/Local switching capability
PP0
No Parallel Poll capability
DC1
Device Clear capability (The SDC and DCL commands can be used.)
DT1
Device Trigger capability (The GET command can be used.)
C0
No Controller capability
E2
Using tri-state bus drivers