Standard,

Fiber Optic and Disposable

Laryngoscope Blades and Handles

Directions
for Use

®

Содержание Satin 4079

Страница 1: ...ADC SatinTM Standard Fiber Optic and Disposable Laryngoscope Blades and Handles Directions for Use ...

Страница 2: ...in this manual identifies a condition or practice which if not corrected or discontinued immediately could lead to patient injury ill ness or death WARNING Federal law restricts this device to sale by or on the order of a physician or li censed healthcare practitioner WARNING ONLY TRAINED PERSONNEL SHOULD USE A LARYNGOSCOPE FOR INTUBATION WARNING ADC recommends that spare lamps batteries and repla...

Страница 3: ...ures ADC Satin Laryngoscope blades are crafted from rugged stainless steel ADC Disposable Fiberoptic Laryngoscope blades are crafted from poly carbonate and are for single use only and should not be cleaned or sterilized Reusable handles are machined from chrome plated brass or anodized aluminum To ensure maximum life and optimum performance the follow ing instructions should be strictly adhered t...

Страница 4: ...ng Basic Cleaning Procedure for Standard and Fiberoptic Blades Prepare a Klenzyme Manu klenz or equivalent cleaning solution as per manufacturer s instructions and soak for ten minutes Rinse devices with lukewarm running tap water for one minute to remove any residual detergent after cleaning Dry with a clean lint free cloth Basic Cleaning Procedures for Standard and Fiberoptic Handles Prepare a K...

Страница 5: ... batteriesmustberemoved Boththeendcapsandhandlesshouldbefullyimmersedinthedisinfectantfora minimumof45minutes Noairbubblesshouldbevisibleontheimmersedcomponentsduringthissoak time Thecomponentsshouldthenberemovedfromthedisinfectantandimmersedinpurifiedwaterto rinseoffalldisinfectant Immersionshouldbeforaminimumofoneminuteandtheprocessshouldbere peatedtwoadditionaltimesusingfreshpurifiedwatereachti...

Страница 6: ... cap on and tighten Figure 3 NOTE Alkaline batteries should be used with your handle for max imum performance and are recommended for replacement Ordi nary carbon zinc batteries may also be used Lamp Replacement Standard Models For correct illumination alignment and assurance of watertight seal during immersion autoclave we recommend using only ADC 4500 1 and 4501 1 replacement lamps for standard ...

Страница 7: ...in use DO NOT attempt to remove this screw from the laryngoscope blade 2 Pull out the green base of the fiberoptic bundle and slide out the light guide Figure 5 Note The fiberoptic bundle is tightly fitted and may be difficult to remove 3 Insert new fiberoptic bundle by reversing the above steps 4 Tighten the locking screw ensuring that the fiberoptic bundle is securely attached to the laryngoscop...

Страница 8: ...sconsin 1 5 Wisconsin 2 Wisconsin 3 Wisconsin 4 C Cell Handle AA Cell Handle Stubby Handle STANDARD MODELS DESCRIPTIONS Standard Replacement Lamp 4500 4500 4500 4501 4501 4500 4500 4500 4501 4501 4501 4500 4500 4500 4501 4501 4501 NA NA NA Standard Blade 4070 4071 4072 4073 4074 4080 0 4080 4081 4082 4083 4084 4090 4091 40915 4092 4093 4094 4065 4066 4067 ...

Страница 9: ... REPLACEMENT PARTS CHART Fiberoptic Light Guide 4070FLP 00 4071FLP 00 4072FLP 00 4073FLP 00 4074FLP 00 NA 4080FLP 00 4081FLP 00 4082FLP 00 4083FLP 00 4084FLP 00 NA NA NA NA NA NA NA NA NA Fiberoptic Replacement Lamp NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA 4532 1 old 5111N 4 new 4532 1 old 5111N 4 new 4532 1 old 5111N 4 new Fiberoptic Blade 4070F 4071F 4072F 4073F 4074F NA 4080F 4081F 40...

Страница 10: ...onsin 0 Wisconsin 1 Wisconsin 1 5 Wisconsin 2 Wisconsin 3 Wisconsin 4 C Cell Handle AA Cell Handle Stubby Handle 10 SPU MODELS DESCRIPTIONS REPLACEMENT PARTS CHART Fiberoptic Blade NA 4071D 20 4072D 20 4073D 20 4074D 20 4080 0D 20 4080D 20 4081D 20 4082D 20 4083D 20 4084D 20 NA NA NA NA NA NA NA 4065D 20 NA ...

Страница 11: ... Incidental special or consequential damages Some states do not allow the exclusion or limitation of incidental special or consequential dam ages so this limitation may not apply to you To Obtain Warranty Service Send item s postage paid to ADC Attn Repair Dept 55 Commerce Dr Hauppauge NY 11788 Please include your name and address phone no proof of purchase and a brief note explaining the problem ...