Otoscope
Coax Ophthalmoscope
Dermascope
Throat Illuminator
Power Handles
Instruction Manual
Страница 1: ...1 Questions Call ADC toll free 1 800 232 2670 Otoscope Coax Ophthalmoscope Dermascope Throat Illuminator Power Handles Instruction Manual ...
Страница 2: ...ument Heads 9 6 Diagnostix Otoscope 9 10 7 Diagnostix Ophthalmoscope 10 12 8 Diagnostix Dermascope 12 13 9 Diagnostix Throat Illuminator 13 10 Replacing the Lamp 13 11 Spare Parts and Accessories 13 12 Cleaning and Disinfection 14 13 Technical Specifications 15 14 Maintenance 15 15 Electromagnetic Compatibility 15 19 16 Warranty 19 17 Quality Standards 20 18 How to Contact Us 20 ...
Страница 3: ...3 DiagnostixTM Otoscope DiagnostixTM Coax Ophthalmoscope DiagnostixTM Dermascope DiagnostixTM Throat Illuminator DiagnostixTM PMV Otoscope DiagnostixTM Coax Plus Ophthalmoscope ...
Страница 4: ...d accessories to ensure safe and functional use of this product 3 WARNINGS AND PRECAUTIONS You have purchased a high quality ADC DiagnostixTM instrument which has been manufactured to the highest global standards and is subjected to rigorous quality control Read these instructions for use carefully before putting the unit into operation and keep them in a safe place If you should have any question...
Страница 5: ...rectedintothepatient seyebe limitedtotheminimumlevelwhichisnecessaryfordiagnosis Infants aphakesandpersons withdiseasedeyeswillbeatgreaterrisk Theriskmayalsobeincreasedifthepersonbeing examinedhashadanyexposurewiththesameinstrumentoranyotherophthalmicinstrument usingavisiblelightsourceduringtheprevious24hours Thiswillapplyparticularlyiftheeye hasbeenexposedtoretinalphotography Warning Otoscope MUS...
Страница 6: ...echargeable Handle 5560 3 5v requires desk charger Plug In Rechargeable Handle 5460 3 5v 230v or 120v Note These handles are compatible with ADC Riester and Welch Allyn 3 5v instrument heads Welch Allyn is a registered trademark Riester is a registered trademark 4 3 Inserting and Removing Rechargeable Batteries Screw off the battery cap on the lower part of the handle Depending on which handle you...
Страница 7: ...ossible light output we recommend always fitting high quality batteries as described in 4 3 If you suspect that liquid or moisture could have entered the handle it must not be charged under any circumstances This could lead to a life threatening electric shock especially in the case of plug in handles To extend the service life of the battery the battery should only be charged once the light inten...
Страница 8: ...ent heads are equipped with a security feature Use enclosed hex key to lock head onto power handle To secure insert hex key into slot on head and tighten by turning clockwise Keep the hex key in a convenient place in case the head must be removed for cleaning or service NOTE Locking the head is recommended on wall mounted systems only 4 6 Adjusting the Light Intensity With the ADC AdtronicTM elect...
Страница 9: ...el Lens for Magnification Standard Series The swivel lens 3x max is fixed to the device and can be swiveled 360 Fig 3 PMVSeries The focusing wheel allows you to adjust the focusing range The adjustment wheel moves up and down to focus the lens Fig 4 6 4 Insertion of External Instruments into the Ear Standard Series only If you wish to insert external instruments into the ear e g tweezers you have ...
Страница 10: ...ection lens can be adjusted on the lens wheel The following correction lenses are available Coax 3 5v Ophthalmoscope Fig 6 Plus 1 10 12 15 20 40 Minus 1 10 15 20 25 30 35 Coax Plus 3 5v Ophthalmoscope Fig 7 Plus 1 45 in single steps Minus 1 44 in single steps The values can be read off in the illuminated field of view Plus values are displayed in green numbers minus values with red numbers 7 3 Ape...
Страница 11: ... a dilated pupil and general examination Coax Coax Plus Karo Grid For topographic determination of retina changes Coax Plus Only Slit To help determine levels of tumors and lesions Coax Coax Plus Fixation star Measuring eccentric fixation or locating lesions Coax Coax Plus 7 4 Changing Filters Using the filter wheel the following filters can be switched for each aperture Coax Coax Plus Ophthalmosc...
Страница 12: ...average service life 15 h Coax Plus 3 5 V ophthalmoscope 3 5v 29 mA average service life 100 000 h 8 DIAGNOSTIXTM DERMASCOPE 8 1 Purpose The Diagnostix Dermascope described in these Operating Instructions is produced for early identification of changes of skin pigmentation malignant melanomas 8 2 Focusing Focus the magnifying glass by rotating the eyepiece ring Fig 10 8 3 Contact Plates Two contac...
Страница 13: ...inator LED 3 5 V 28 mA mean life span 100 000h 10 REPLACING THE LAMP All Instrument Heads Remove the instrument head off the battery handle The lamp is located at the base of the instrument head Pull the lamp out of the instrument head with thumb and forefinger or a suitable tool Insert a new lamp CAUTION The pin on the ophthalmoscope lamp must be inserted into the guide groove on the instrument h...
Страница 14: ...ection The instrument heads and handles can be cleaned externally with a moist cloth until visually clean Wipe disinfection as specified by the disinfectant manufacturer Only disinfectants with proven efficacy should be used taking into account the national requirements After disinfection wipe the instrument down with a moist cloth to remove possible disinfectant residues The components that come ...
Страница 15: ...ation equipment can influence medical electrical equipment This ME device is intended for operation in an electromagnetic environment as specified below The user of the device should ensure that it is operated in such an environment The ME device must not be used directly next to or arranged in a stack with other devices If the device has to be operated near to or in a stacked arrangement with oth...
Страница 16: ...he DiagnostixTM instrument employs HF energy solely for an internal function Its HF emission is therefore very low and it is unlikely that neighboring electronic devices will be affected by interference HF emissions according to CISPR 11 Class B Harmonics emissions according to EC61000 3 2 Not applicable The DiagnostixTM instrument is intended for use in all facilities including living quarters an...
Страница 17: ...d output lines 3 A m 3 A m Not applicable The quality of the supply voltage should correspond to that of a typical business or hospital environment Surges IEC61000 4 5 1 kV voltage phase to phase 2 kV voltage phase to earth Not applicable The quality of the supply voltage should correspond to that of a typical business or hospital environment Voltage dips short time interruptions and fluctuations ...
Страница 18: ...radio transmitters should be less than the compliance level at all frequencies as verified by an on site testa Interference is possible in the vicinity of equipment marked with the following symbol Note 1 At 80 MHz and 800 MHz the higher value applies Note2 Theseguidelinesmaynotapplyinallsituations Thepropagationofelectromagneticwavesisinfluencedby reflection and absorption by buildings objects an...
Страница 19: ...2 5GHz For transmitters whose nominal power is not indicated in the table above the distance can be determined using the equation belonging to the respective column where P is the nominal power of the transmitter in Watts W as specified by the manufacturer of the transmitter Note 1 At 80 MHz and 1400 MHz the distance for the higher frequency range applies Note2 Theseguidelinesmaynotapplyinallsitua...
Страница 20: ...QUALITY STANDARDS Device standard This device is manufactured to meet the European and United States standards for ISO 62471 2008 ISO 15004 ISO 10942 ISO 15004 1 and ISO 15004 2 Electromagnetic compatibility Device fulfills the stipulations of the International standard IEC60601 1 2 18 HOW TO CONTACT US To register your product and obtain further detailed user information about our products and se...