background image

STOP

if you meet any resistance. The sensing unit is on the very tip of the probe

and there is no need to insert the probe deep into the rectum.

Dispose of the used probe cover and wash the thermometer as recommended. The

peak temperature should be reached in approximately 25-35 seconds.

Axillary Use: 

Wipe armpit with a dry towel. Place the probe 

in the patient’s armpit (Figure 3) and keep the patient’s arm

pressed firmly against the body. In order to achieve a better 

axillary reading, a minimum measuring time of 90 seconds or

longer is recommended, depending on placement of the probe

and closure of the arm against the body.

Battery Replacement

1.   When the symbol “      ” appears, the battery  needs to

     be replaced.

2.   Remove battery compartment lid (Figure 4).

3.   Use a pointed object such as a toothpick to pry loose the

     old battery. Avoid using any sharp metal object.

4.   Place new 1.55V coin-type battery (LR41 or equivalent) in the chamber with the 

     positive (+) side facing up or towards the display of the unit.

5.   Replace the battery compartment lid.

Cleaning and Disinfecting

       

Do not use corrosive detergent to clean the unit.

       

Clean and disinfect the thermometer before and after each use.

Cleaning Process

1.  Wipe probe with a cloth dipped in a disinfectant solution such as 70% 

     isopropyl (rubbing) alcohol or soap and water.

2.  Rinse off disinfectant residue.

Care and Storage

•   Do NOT drop or crush; this device is not shock proof.

•   Do NOT dismantle or modify device other than replacing the battery.

•   Do NOT store the device in an extreme environment with direct sunlight or high

temperatures.

•   Store the unit in the protective case while not in use.

Figure 3

Figure 4

BATTERY

Specifications

   

Size:                            

5”L x 3/4”W x 3/8”H  (12.7cm L x 1.9 cm W x .95 cm H)

   

Weight:                        

0.388 oz.                      (11 grams)

   

Display:                       

LCD, Liquid Crystal Display

   

Battery:                       

1.55V (LR41) 

   

Battery Life:                

Estimated 4,500 measurements

   

Measurement Time:    

Oral 30* seconds / Rectal 25-35 seconds / Axillary 90 seconds

   

Measurement Range:   

90.0°F to 109.9°F  (32.0°C to 42.9°C)

                                          96.0°F to 107.0°F  

±

0.2°F / 1°C                   (34.0°C to 42.0°C)

                                             at a room temperature of 64.4°F to 82.4°F     (18.0°C to 28.0°C)

   

Accuracy:

                     90.0°F to 96.0°F      

±

0.4°F / 0.2°C              (32.0°C to 34.0°C)

                                             and 107.0°F to 109.9°F                               (42.0°C to 42.9°C)

   

Measurement mode:   

Real measurement with peak-hold method

   

Audible Alarm:             

1 beep at start-up. Continuous beep at measurement completion

   

Auto Turn-off               

In approximately 8 ±1 minutes

   

Memory:                      

Last reading

   

Operating Conditions:

  

50°F to 104°F  (10°C to 40°C) 

                                          Humidity of 15% to 95% (non-condensing)

   

Storage Conditions:

    

-13°F to 131°F  (-25°C to 55°C) 

                                          Humidity of 15% to 95% (non-condensing)

   

Standards:                   

Complies with EN 12470-3, ASTM E 1112-00,

                                       IEC 60601-1, IEC 60601-1- 2, IEC 60601-1-11

                                       Electromagnetic compatibility: device fulfills the stipulations

                                       of standard IEC 60601-1-2
                                       

The manufacturer recommends a calibration check every 2 years.

                                       

*30-second response time achieved in water bath test.

LIMITED WARRANTY

ADC warrants its products against defects in materials and workmanship under normal house-

hold use and service as follows: 

1. Warranty service extends to the original retail purchaser only and commences with the date

    of delivery.

2. Your Adtemp Digital Stick Thermometer is warranted for two years from date of purchase.

What Is Covered: 

Replacement of parts, and labor.

What Is Not Covered: 

Transportation charges to ADC. Batteries where supplied. Damages

caused by abuse, misuse, accident, or negligence, incidental, special, or consequential 

damages. Some states do not allow the exclusion or limitation of incidental, special, or 

consequential damages, so this limitation may not apply to you.

To Obtain Warranty Service: 

Send item(s) postage paid to ADC, Attn: Repair Dept., 

55 Commerce Dr., Hauppauge, NY 11788. Please include your name and address, phone no., proof

of purchase, and a brief note explaining the problem.

Implied Warranty: 

Any implied warranty shall be limited in duration to the terms of this 

warranty and in no case beyond the original selling price (except where prohibited by law).

This warranty gives you specific legal rights and you may have other rights which vary from state to state.

For Australian Consumers: 

Our goods come with guarantees that cannot be excluded under

the Australian Consumer Law. You are entitled to a replacement or refund for a major failure and

compensation for any other reasonable foreseeable loss or damage. You are also entitled to

have the goods repaired or replaced if the goods fail to be of acceptable quality and the failure

does not amount to a major failure.

IB p/n 93-413-00 rev 8  (7/20/21)

Printed in China

LATEX

Microlife Corporation
9F, 431 RuiGuang Road
NeiHu, Taipei 114, Taiwan, R.O.C.

Medical Device Safety Service GmbH
(MDSS GmbH)
Schiffgraben 41, 30175 Hannover, GERMANY

Mfg. by: ADC

®

55 Commerce Drive, Hauppauge, NY 11788

Inspected in the USA

Made in China

Assembled in Vietnam

tel: 631-273-9600

toll free: 1-800-232-2670

fax: 631-273-9659

www.adctoday.com

email: [email protected]

1639

EC REP

Отзывы: