2.5V EENT Instruments
Instructions for Use
LIMITED WARRANTY
American Diagnostic Corporation (ADC
®
) warrants its products against defects in materials and
workmanship under normal use and service as follows:
1. Warranty service extends to the original retail purchaser only and commences with
the date of delivery.
2. The entire system is warranted for one year.
What Is Covered:
Repair, or replacement of parts, and labor.
What Is Not Covered:
Transportation charges to ADC. Batteries where supplied. Damages caused by abuse, misuse,
accident, or negligence. Incidental, special, or consequential damages. Some states do
not allow the exclusion or limitation of incidental, special, or consequential damages,
so this limitation may not apply to you.
To Obtain Warranty Service
:
Send item(s) postage paid to ADC, Attn: Repair Dept., 55 Commerce Dr., Hauppauge,
NY 11788. Please include your name and address, phone no., proof of purchase,
and a brief note explaining the problem.
Implied Warranty:
Any implied warranty shall be limited in duration to the terms of this warranty and in
no case beyond the original selling price (except where prohibited by law).
This warranty gives you specific legal rights and you may have other rights which vary
from state to state.
For Australian Consumers:
Our goods come with guarantees that cannot be excluded under the Australian Consumer Law.
You are entitled to a replacement or refund for a major failure and compensation for any other
reasonably foreseeable loss or damage. You are also entitled to have the goods repaired or
replaced if the goods fail to be of acceptable quality and the failure does not amount to a major failure.
For our European Customers
On request, send to ADC by e-mail
)
, we can send to you this manual on
paper form within 7 calendar days at no additional cost to the user.
Our website,
https://www.adctoday.com
,
where these instructions for use are available
fulfills the requirements of personal data protection, according to Directive 95/46/EC and
GDPR on the protection of individuals with regard to the processing of personal data and on
the free movement of such data.
Any serious incident that has occurred in relation to this medical device should be reported to
ADC and the competent authority of the Member State in which the user and/or patient is
established.
Symbol Definitions
The following symbols are associated with your diagnostic instrument.
Environment
Transportation / Storage:
-4°F to 120.2°F (-20°C to 49°C)
Relative Humidity: 95 %
500 hPa – 1060 hPa, Altitude
Environmental Operating
Range:
50°F to 120.2°F (10°C to 49°C)
Relative Humidity: 95 % (Max)
500 hPa – 1060 hPa, Altitude
Maximum Operating
Temperature:
95°F (35°C)
Troubleshooting
The following guidelines should be used to determine if your device has reached its end of life or
requires servicing. If the corrective actions described in this section do not resolve your issues, please
see the warranty section of this manual to have your device serviced.
Otoscope and Accessories Troubleshooting Guide
Trouble Area Possible Cause Corrective Action
Lamp burned Replace the lamp with
out ADC part #5211-4
No light output Wrong lamp Replace the lamp with
installed ADC part #5211-4
Batteries are Replace with fresh
depleted “C” batteries
Instrument head Thread/connection Service
will is required
not connect damaged
Wrong specula Use only Welch Allyn brand
Specula will not used or compatible
attach properly
Specula mount is damaged Service required
Loose or misaligned Lens mount Service
magnifying lens is damaged required
Lens is dirty Clean with lint free cloth
Poor or obstructed view
through lens Lens is damaged Service required
Lamp near end Replace the lamp with
Dim light of useful life ADC part #5211-4
output
Batteries are Replace with fresh
depleted “C” batteries
Visible corrosion on Damage from excessive Service
instrument or accessories moisture required
Ophthalmoscope Troubleshooting Guide
Trouble Area Possible Cause Corrective Action
Aperture dial is Rotate aperture dial
between positions
Lamp burned Replace the lamp with
No light output out ADC part #5210-4
Wrong lamp Replace the lamp with
Installed ADC part #5210-4
Instrument head Lamp not fully inserted Re-insert lamp
will
not connect Thread/connection is damaged Service required
Spot is not The aperture dial is not centered Move aperture dial to
centered the full detent position
Lamp near end of useful life Replace the lamp with
Dim light ADC part #5210-4
output
Batteries are depleted Replace with fresh “C” batteries
Aperture or lens wheel will not Dirt in or damage Service
turn or does so with difficulty to mechanism required
Visible corrosion on instrument Damage from excessive moisture Service required
Bent Arm Illuminator and Accessories Troubleshooting Guide
Trouble Area Possible Cause Corrective Action
Lamp burned out Replace the lamp with
ADC part #5211-4
No light
output Wrong lamp installed Replace the lamp with
ADC part #5211-4
Batteries are depleted Replace with fresh “C” batteries
Accessories do not attach Accessories may have Obtain replacement
securely or operate properly reached end of useful life accessories
Instrument head will Thread/connection Service
not connect is damaged required
Lamp near end of Replace the lamp with
Dim light useful life ADC part #5211-4
output
Batteries are depleted Replace with fresh “C” batteries
Visible corrosion on Damage from excessive Service
instrument moisture required
ˆ
ˆ
Symbol
Definition
Do not dispose of this product
as unsorted municipal waste.
Prepare this product for reuse or separate
collection as specified by Directive
2002/96/EC of the European Parliament
and the Council of the European Union
on Waste Electronic and Electrical
Equipment (WEEE). If this product is
contaminated, this directive does not apply.
Date of manufacture
Batch Code
Medical Device
Unique device identifier
Non-sterile
Do not use if package is damaged
Importer
Distributor
Catalog Number
Symbol
Definition
Important Caution
Not made with natural rubber latex
Phthalate free
Consult instructions for use
Meets the general safety and
performance requirements of
Regulation (EU) 2017/745 of the
European Union
Authorized Representative in the
European Community/European Union
Manufacturer
Follow Instructions for Use
TYPE B.
Indicates this is a product
with Type B applied parts
Keep dry
Circumference size
93-5215-00 rev 11.qxp_Layout 1 12/1/22 3:53 PM Page 1