![Adaptica 2WIN-S Скачать руководство пользователя страница 20](http://html.mh-extra.com/html/adaptica/2win-s/2win-s_operators-manual_2846936020.webp)
20
Patient age
2WIN-S uses the patient
’
s birthdate / age range as a corrective parameter to determine the refraction. This
information is used in particular to compensate accommodation in young patients with hyperopia-based
refraction error. Take care of selecting the right birthdate / age range before taking the measurement. Take
into consideration that the default age range is 0-2 years.
Control Panel functionalities
Button
Power
1 – Measure
2 – Connection
State
Flashing Green
The system is booting
up
Exam ongoing
-
Fixed Green
The system has
successfully started.
There is a blinking
when the activity is
normal
Measure well done
2WIN-S correctly
connected to the
application
Fixed Red
Error: please shut down
the system and power it
on again
Measure error:
repeat the
measurement
2WIN-S not
connected: check
connections’ settings
Flashing
Yellow
The system is shutting
down
-
Connection ongoing
(Wi-Fi/BT)
Fixed Yellow
Firmware updating
-
2WIN-S not
connected to the
application
Flashing blue
Pairing ongoing with
application
Button
Power
1 – Measure
2 – Connection
Function
Short press
System power up
-
-
Long press
2 sec: system shut
down
More: force power off
2 sec: start/stop a
measurement; the
exam is
automatically saved
2 sec, until flashing
yellow: WiFi\BT
connection reset
4 sec, until flashing
blue: start application
pairing
9.2 Preparation of the patient
The operator shall instruct the patient to look inside the tube and to try to open wide his eyes so to avoid
eyelashes covering the pupils.
The operator shall instruct the patient to look, during the measurement, at the fixation target positioned at
the bottom of the device: three white dots in the shape of a triangle. Most importantly, the patient must not
follow with his eyes the moving red lights that turn on during the exam but concentrate on the fixation target.
In case of monocular measurement, ask the patient to cover his/her other eye with the provided occluder.