Adam Equipment 

 
 

 
 

MDW MCW 

 Service Manual 

 

                                         

Part No- 700660171 – rev C1 January 2012 

 

 

 
 

Содержание MCW

Страница 1: ...Adam Equipment Company Ltd 2012 Adam Equipment MDW MCW Service Manual Part No 700660171 rev C1 January 2012 ...

Страница 2: ...Adam Equipment Company Ltd 2012 ...

Страница 3: ...RIPTION 2 2 0 SPECIFICATION TABLE 3 3 0 TROUBLE SHOOTING GUIDE 4 4 0 ERROR MESSAGES FOR MDW MCW 4 5 0 LOAD CELL REMOVAL AND REPLACEMENT 6 6 0 LOAD CELL WIRE TO INDICATOR CONNECTIONS 8 7 0 CALIBRATION FOR L MODELS 11 8 0 PARAMETERS 13 9 0 WARRANTY INFORMATION 20 ...

Страница 4: ... and 300kg x 0 05kg as well as calculating your BMI following height entry into the Indicator The MCW 300L is the wheelchair version and has the same functions and specifications as the MDW 300L A single load cell is used for all 3 L models mounted to the weighing platform and connected to a digital Indicator through a pillar via a cable The digital Indicator used for both capacity models is very ...

Страница 5: ...t Company Ltd 2012 MODEL CAPACITY AND DIVISION BASE SIZE mm MDW 160M 160kg x 0 1kg 275 x 375 MDW 250L 250kg x 0 1kg 275 x 375 MDW 300L 300kg x 0 05kg 275 x 375 MCW 300L 300kg x 0 05kg 450 x 450 2 0 SPECIFICATION TABLE ...

Страница 6: ..._ _ 0 Zero point exceeds range Load cell damage main PCB fault Weight left on weighing platform 0_ _ _ _ _ _ Zero point is under range Load cell damage main PCB fault Load cell connection problem _ _ _ _ _ Ad A d exceeds the maximum range Improper calibration of the indicator Load cell damage main PCB fault Ad_ _ _ _ _ A d is below the minimum range Improper calibration of the indicator Load cell ...

Страница 7: ...ell the next step is to measure the output of the load cell this can be done by placing the 2 probes between the sig and sig wires In order to obtain an accurate reading the sig and sig wires must be detached de soldered from the connector You should expect to see a positive voltage reading of between 0 and 2 milli volts when the load cell assembly is at no load and when a downward force is applie...

Страница 8: ...emove then unscrew the 4 x fixing screws that secure the chair assembly to the main frame This then exposes the weighing platform for both the MCW L and MDW L Unscrew the 4 x load cell fixing bolts that secure the load cell to the weighing platform and remove Turn the platform over and unscrew the 4 x load cell fixing bolts that secure the load cell to the main base The load cell can then be remov...

Страница 9: ...7 Adam Equipment Company Ltd 2012 MCW MCW and MDW L ...

Страница 10: ...ost effective method of repair If this is the case the PCB number and model of machine should be made clear so that the PCB can be programmed and checked first once checked the PCB can then sent to be fitted directly to the Indicator Calibration can then be performed and the unit should function correctly All load cell to Indicator cables are 4 core and wired via a connector following the pin numb...

Страница 11: ...oise weight should be adjusted individually using its own beam The zero poise weight should only be adjusted so that a zero point can easily be achieved with the zero screw this poise weight has no direct effect on the calibration of the MDW The knife edges in the base need to be located into the shackles that contains the bearings if positioned correctly the top pan should always swing freely whe...

Страница 12: ...10 Adam Equipment Company Ltd 2012 ...

Страница 13: ...e UNIT key to select the calibration weight unit that you want to calibrate in Press the Print Hold key to choose a different calibration weight value 50kg 100kg 150kg 200kg 250kg 300kg or 100lb 200lb 300lb 400lb 500lb 600lb Then put on the weight that you selected and press the TARE key to confirm the chosen standard weight that was selected The displayed data will flash on the display and if the...

Страница 14: ...not correct repeat the calibration process again A FACTORY Calibration can be performed if you have problems with the USER calibration to carry out a FACTORY Calibration press and hold down simultaneously the UNIT and TARE buttons for 3 seconds whilst the reading is at ZERO from this point on follow the procedure above from point 2 ...

Страница 15: ...arameters setting summary 8 0 PARAMETERS Parameter Available Settings Factory Setting Setting A o t 00 15 05 Auto off time No auto off 00 01 15 minutes auto off time P H 0 1 2 1 Print and Hold Button function 0 Only Print Function 1 Only Hold Function 2 both HOLD and PRINT function To PRINT push button once To HOLD push button down for 3 seconds H t 0 4 0 Hold time 0 no time limit 1 10 seconds 2 3...

Страница 16: ...VOT 7 7 00 1 0 0130 CONNECTING ROD LEVER 8 7 00 1 2 0002 BRACKET 9 7 00 2 3 0004 PLASTIC COVER 10 7 00 1 2 0003 EYE LOOP 11 7 00 1 0 0137 PILLAR HEAD T SHAPE 12 7 00 2 0 0047 HEIGHT ROD 13 7 00 1 0 0129 BASE TO INDICATOR CONNECTING ROD 14 7 00 1 0 0142 PILLAR 15 7 00 2 0 0046 PLASTIC PLATFORM COVER 16 7 00 1 0 0133 METAL PLATFORM 17 7 00 1 0 0139 CENTRE LINK 18 7 00 1 0 0141 LONG LEVER 19 7 00 1 0...

Страница 17: ...15 Adam Equipment Company Ltd 2012 ...

Страница 18: ...FASCIA 7 7 00 4 0 0037 MAIN DISPLAY PCB STATE WHICH REVISION 8 7 00 1 0 0147 LOAD CELL LCT LAE A 350kg cell 9 7 00 1 0 0148 PILLAR 10 7 00 1 2 0004 PILLAR HOLDER 11 7 00 4 0 0073 ON OFF SWITCH 12 7 00 4 0 0074 POWER IN SOCKET 13 7 00 3 0 0012 COMPLETE PACKAGING SET 3 02 4 0 9156 12v DC 800ma ADAPTOR UK 3 02 4 0 9160 12v DC 800ma ADAPTOR USA 3 02 4 0 9159 12v DC 800ma ADAPTOR OZ 3 02 4 0 9157 12v D...

Страница 19: ... PILLAR 18 7 00 1 0 0259 MCW 300L HANDLE 19 7 00 1 0 0243 MCW 300L METAL WEIGHING PLATE 20 7 00 2 0 0076 MCW 300L WHEEL WITH BRAKE REAR 21 7 00 1 0 0260 MCW 300L WHEEL SWIVEL FRONT 22 7 00 2 0 0073 MCW 300L FOOT REST FRAME ASSEMBLY 23 7 00 1 0 0261 MCW 300L WHEEL FRAME EXC WHEELS 24 7 00 1 0 0241 MCW 300L CHAIR FRAME ASSEMBLY 25 7 00 1 0 0262 MCW 300L ARM REST METAL PART x 2 26 7 00 3 0 0019 MCW 3...

Страница 20: ...18 Adam Equipment Company Ltd 2012 ...

Страница 21: ...19 Adam Equipment Company Ltd 2012 ...

Страница 22: ...orrect documentation for a claim to be processed All claims are at the sole discretion of Adam Equipment This warranty does not cover equipment where defects or poor performance is due to misuse accidental damage exposure to radioactive or corrosive materials negligence faulty installation unauthorised modifications or attempted repair or failure to observe the requirements and recommendations as ...

Страница 23: ...t in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense Shielded interconnect cables must be employed with this equipment to insure compliance with the pertinent RF emission limits governing this device Changes or modifications not expressly approved by Adam Equipment could void the user s authority to o...

Страница 24: ...s Health and Fitness Scales Retail Scales for Price computing For a complete listing of all Adam products visit our website at www adamequipment com Copyright by Adam Equipment Co Ltd All rights reserved No part of this publication may be reprinted or translated in any form or by any means without the prior permission of Adam Equipment Adam Equipment reserves the right to make changes to the techn...