© Adam Equipment Company 2012 

11

 

 

 

5.0  USER SETTINGS 

 
The grain scale can be set up by the USER to enable or disable container sizes, 
units of yield and dockage results in order to make operation and recalling of 
results faster. 
 
The scale has 3 settings available. 

 

FUNCTION 

SECTION 

DESCRIPTION 

Yield  

See section 5.2 

Selects which units of yield are to be displayed  

Container size 

See section 5.3 

Selects the container sizes available to be used 

Dockage 

See section 5.4 

Selects the type of dockage to be displayed in the 
test results 

 
 

5.1 

USER SETTING ENTRY 

 

  Press the 

[ON/OFF] key

 to turn the balance off. 

 

  Press and Hold the 

[CON/UNIT]

 key and press the 

[ON/OFF]

 key to switch 

the scale on, the scale will show “grn-3.3 GRN” and then enter the USER 
setting mode. The display will show "X k.hl" which is the first parameter. 

 

5.2  

YIELD 

Yield can be show n in Kilograms per hectolitre (k.hl) and/or Pounds per bushel (L.bu).  
 

1) The display w ill show  ‘‘ X k.hl’’  (X being the current setting) 

 

2)  Press 

[TARE]

 to set ON or OFF for k.hl 

 

3)  Press 

[STORE]

 to select L.bu 

 

4)  Press 

[TARE]

 to set ON or OFF for L.bu 

 

5)  Press 

[RECALL/CE]

 t o confirm settings 

 

Содержание CQT1752GR

Страница 1: ... Adam Equipment Company 2012 Adam Equipment Grain Scale Model CQT1752GR Part No 3086611514 Rev A1 April 2012 Software Ver 3 3 ...

Страница 2: ... Adam Equipment Company 2012 ...

Страница 3: ...N THE SCALE 6 4 2 ZEROING TARE 6 4 3 WEIGHING 6 4 4 WEIGHING UNITS 6 4 5 GRAIN DETERMINING 7 5 0 USER SETTINGS 11 5 1 USER SETTING ENTRY 11 5 2 YIELD 11 5 3 CONTAINER SIZE 12 5 4 DOCKAGE 12 6 0 PARAMETERS 13 6 1 PARAMETER SETTING 13 6 2 FILTER 13 6 3 ZERO TRACKING 13 6 4 STABILITY 14 6 5 AUTO POWER OFF 14 6 6 SETTING THE BACKLIGHT 14 7 0 CALIBRATION 15 8 0 TROUBLE SHOOTING 16 8 1 ERROR MESSAGES 16...

Страница 4: ...ample weights 3 0 SET UP 3 1 UNPACKING AND SETTING UP YOUR SCALE 1 Remove the shipping protection screw and round cover as shown below using the hex allen key supplied in the box It is advisable to keep the shipping protection screw and the round cover for shipping to another location in the future Shipping protection screw Round cover protecting the load cell 2 Place the pan which comes separatel...

Страница 5: ...ff key Turns the balance on off Con Unit key Selects the container size and Unit switches between weighing modes and results Tare key Tares the balance Store key Stores the initial sample and subsequent weights following screening Recall CE key Recalls stored weighing results and CE clears previously stored results from the memory ...

Страница 6: ...y compartment cover and insert six batteries R6P LR6 AA size into the battery compartment Battery compartment screws Battery compartment cover Batteries are not included ZERO indicator Indicates when the scale is at zero STABLE indicator Indicates when the reading is stable ...

Страница 7: ...ower sources Do not operate near large users of electricity such as welding equipment or large motors Do not leave the batteries inside the scale if you are not using it for a long time Keep free from vibration Do not place near heavy or vibrating machinery Avoid high humidity that might cause condensation and keep away from direct contact with water Do not place near open windows air conditioning...

Страница 8: ... set a new zero point Zero will be set if the reading on the scale is less than 70g This may be necessary if the weight is not reading zero with no weight on the pan The zero indicator will show in the top left corner of the LCD If you are using a container to weigh then you can place this on the platform and press the Tare key the display will show zero and NET on the display You can then weigh a...

Страница 9: ... with the grain The screen shows the net weight the first net weight of the grain in this example the total weight is 400g qt pt Smp 0 5lt lt stored At this point you can press the CON UNIT key to view the yield values calculated in k hl and L bu based on the net weight in the container 4 To store this weight as the initial sample value press the STORE key the display will show Store smp and then ...

Страница 10: ...STORE to be performed when the container is placed back onto the top pan The first dockage result will then be displayed and by pressing the CON UNIT key further results can be viewed including dockage the net weight in grams which was used in the calculation and also the new yield values Pressing the RECALL CE key unlocks the display for further weighings Pressing the RECALL CE key again continuo...

Страница 11: ...e displayed Negative dockage showing the of grain removed after screening and indicating the 1st recorded set of test results qt pt Smp 0 5lt lt stored Positive dockage showing the of grain remaining after screening and indicating the 1st recorded set of test results qt pt Smp 0 5lt lt stored Yield in Kilograms per hectolitre of the remaining grain following the first screening qt pt Smp 0 5lt lt ...

Страница 12: ...CALL CE key for 3 seconds A message of CLEAR will be displayed to confirm deletion of the stored values and the indicator above Smp stored will go out New tests can now be performed If the balance is turned off during the weighing process it will retain the stored information in the memory so that you could then complete the tests without having to start again due to a loss of power Once a sample ...

Страница 13: ...ee section 5 4 Selects the type of dockage to be displayed in the test results 5 1 USER SETTING ENTRY Press the ON OFF key to turn the balance off Press and Hold the CON UNIT key and press the ON OFF key to switch the scale on the scale will show grn 3 3 GRN and then enter the USER setting mode The display will show X k hl which is the first parameter 5 2 YIELD Yield can be shown in Kilograms per ...

Страница 14: ...et ON or OFF for 1lt 7 Press STORE to select 0 5lt 8 Press TARE to set ON or OFF for 0 5lt 9 Press RECALL CE to confirm settings 5 4 DOCKAGE The dockage results can be set to display negative and or positive dockage 1 The display will show X X being the current setting 2 Press TARE to set ON or OFF for 3 Press STORE to select positive 4 Press TARE to set ON or OFF for positive 5 Press RECALL CE to...

Страница 15: ...tion allows the USER to adjust the speed of display update a setting of 1 will mean a faster response on the display from when the weight is placed on the pan default is 2 1 The display will show X FIL X being the current setting 2 Press the TARE key to increment through the various settings 1 6 3 Press the RECALL CE key to confirm the setting chosen 6 3 ZERO TRACKING The zero tracking function in...

Страница 16: ...er will switch off 20 PWR 20 seconds before the power will switch off 30 PWR 30 seconds before the power will switch off 40 PWR 40 seconds before the power will switch off 50 PWR 50 seconds before the power will switch off 60 PWR 60 seconds before the power will switch off 70 PWR 70 seconds before the power will switch off 80 PWR 80 seconds before the power will switch off 90 PWR 90 seconds before...

Страница 17: ...ess the TARE and STORE keys together then release these keys 2 The display will show UnLOAD remove any weight from the pan 3 Press the TARE key the weight value used for the last calibration will be displayed 4 To select a different calibration weight value press the STORE key This will cycle you through the weight values that can be selected for your calibration values available are 500g 1000g an...

Страница 18: ... high Weight on the pan when turning on Improper calibration of the scale Damaged load cell Damaged Electronics Remove any weight from the stainless steel pan Recalibrate ADL A D Value too low Pan is not installed when turning on Calibration not correct Damaged load cell Install pan and power on Recalibrate Low battery indicator Batteries may be flat Change the batteries FAIL Calibration failure I...

Страница 19: ...31 MAIN PCB inc DISPLAY PCB 3 08 2 3 2003 UPPER HOUSING 3 08 2 3 2004 LOWER HOUSING 7 00 4 0 0023 POWER SUPPLY STANDARD MODEL 12Vac 150ma UK 7 00 4 0 0024 POWER SUPPLY STANDARD MODEL 12Vac 150ma USA 7 00 4 0 0025 POWER SUPPLY STANDARD MODEL 12Vac 150ma OZ 7 00 4 0 0026 POWER SUPPLY STANDARD MODEL 12Vac 150ma EURO 7 00 4 0 0027 POWER SUPPLY STANDARD MODEL 12Vac 150ma SA 3 08 1 0 2043 CARRY CASE 3 0...

Страница 20: ...hester Stabilisation Time 2 3 seconds Operating Temperature 0ºC to 40ºC 32ºF to 104ºF Power Supply 12VAC 150mA adapter or 6x AA batteries Calibration Push button calibration using external mass Calibration Mass 500g 1000g or 1500g user selectable Display 6 digit LCD 18mm digits with backlight Balance Housing ABS Plastic Pan Size 145mm ø 5 7 ø Overall Dimensions w x d x h 173 x 255 x 86mm 6 8 x 10 ...

Страница 21: ...inal packaging and with correct documentation for a claim to be processed All claims are at the sole discretion of Adam Equipment This warranty does not cover equipment where defects or poor performance is due to misuse accidental damage exposure to radioactive or corrosive materials negligence faulty installation unauthorised modifications or attempted repair or failure to observe the requirement...

Страница 22: ...in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense Shielded interconnect cables must be employed with this equipment to insure compliance with the pertinent RF emission limits governing this device Changes or modifications not expressly approved by Adam Equipment could void the user s authority to ope...

Страница 23: ... Adam Equipment Company 2012 ...

Страница 24: ...Health and Fitness Scales Retail Scales for Price computing For a complete listing of all Adam products visit our website at www adamequipment com Copyright by Adam Equipment Co Ltd All rights reserved No part of this publication may be reprinted or translated in any form or by any means without the prior permission of Adam Equipment Adam Equipment reserves the right to make changes to the technol...

Отзывы: