The display will show the parameter number then the word describing the function.
Press the
[Tare]
key to enter a parameter.
Press the
[Zero]
key to exit the factory parameter section. Display will show “
JP on
”.
Remove the jumper. The scale will return to weighing.
When a parameter is entered by pressing the
[Tare]
key, the display will guide you
through the parameter selected and the options available.
The parameters available are:
F1
CAL
To enter the Calibration
F2
rES
Resolution selection, not valid for approved units
F3
CAP
Select Capacity
F4
int
Initial Zero Range
F5
rE Z
Re-Zero Range
F6
SCS
Successive Tare Enable
F7
Cnt
Display the A/D counts
F8
ZE M
Set zero method
F9
Ld M
Set low voltage detect mode
8.1
F1
-CALIBRATION
“
F1 CAL
”
Press
[Tare]
“
U
n
LoA
d
”
Empty the pan by removing the load, if there is any
and press
[Tare]
“
L
d1
XX
kg
”
Load the first calibration weight and press
[Tare]
“
L
d2
XX
kg
”
Load the second calibration weight and press
[Tare]
“
PASS
”
Calibration is complete. Remove the calibration weight.
Or, “
FA
i
L L
”
This means calibration has failed. Remove the
calibration weight and repeat the process.
“
JP
on
”
Remove the jumper or shorting of the pins whichever
is used. The scale will return to normal weighing.