19
The average life of this medical device with normal use and in accord with the present user manual,
has been calculated to be 5 years. According with EN62353 we recommend users verify every 2 years
from the date of the first use: dispersion current versus ground, in the box and in the patient
.
MOTOR RUBBER O-RING:
-
If necessary, substitute the “Rubber O-Ring”
LUBRICATING:
-
Do not oil or lubricate the micro motor.
Protect motor from excess oil draining from handpiece.
After lubricating and before autoclaving, stand handpiece by its base on a paper towel and allow
excess oil to drain.
-
The internal gears of the peristaltic pump are self lubricating.
-
Do apply some oil drops in the cylinders of the peristaltic pump rotor after a long period of non-use or
when the pump doesn’t run continuously.
11.3 FUSE REPLACEMENT
Your device is protected by one fuse. The replacement procedure is as follows:
-
ONCE UNIT IS UNPLUGGED
-Remove the fuse holder.
-
Extract the fuse holder, replace the fuse with one of the same type and value 1x4A T (Dim. 5x20).
Note
:
All the information necessary for the repair of faulty components is made available by the manufacturer to the
technical personnel of the SATELEC approved dealer network who carry out the work.
The I-SURGE must be used in accordance with these Instructions for Use.
The I-SURGE has no components which can be repaired by the user. Assembly, modifications
or repairs must only be undertaken by an authorized service organization.
Unauthorized opening of the equipment invalidates all claims under warranty and any other
claims.
Use on electrical circuit complying with the terms of regulation CEI 64-8 section 710.
11.4 OPERATION FAULTS
Fault detected: The central unit does not switch on.
Possible cause
Solution
•Power cord not connected properly
Make sure the power cord is properly plugged into the electrical outlet and the
connector on the back of the unit.
•Unit has not been switched on.
Check the back of the unit to see that the On button has been pressed.
Содержание I-Surge
Страница 2: ...2...
Страница 212: ...212 I I SURGE 40 000 RPM E Type ISO 3964 NO DROP SATELEC II Federal Law I SURGE...
Страница 213: ...213 3...
Страница 214: ...214 5 BY DENTAL SATELEC Repair www acteongroup com E mail satelec acteongroup com...
Страница 217: ...217 1 2 3 1 3 4 5 6 150 mm 5 O 4 4 I SURGE X V I SURGE VI I SURGE SATELEC VII 7 1 1 5 0 1 2 3 ON OFF...
Страница 220: ...220 4 VIII I SURGE BY DENTAL SATELEC I SURGE 8 2 10 8 E type ISO 3964 5 2 11 I SURGE IN OUT 10...
Страница 221: ...221 I I SURGE IX I SURGE I SURGE O X I SURGE...
Страница 222: ...222 6 30 EN 13060 134 C 273 F 5 2 1 2 2 bars 30 88 psi 5 C 65 C 41 F 149 F RH 20 95 10 1 SATELEC...
Страница 223: ...223 10 2 10 3 10 4 I SURGE I SURGE 10 5 B 134 C 273 F 2 1 2 2 bar 5 NaCl 0 9 134 C...
Страница 225: ...225 11 3 1x4A T 5x20 SATELEC I SURGE I SURGE CEI 64 8 710...
Страница 226: ...226 11 4 On PCB SATELEC PCB SATELEC SATELEC ON OFF SATELEC...
Страница 227: ...227 FWD REV SATELEC Possible cause Solution PCB SATELEC SATELEC SATELEC ON OFF SATELEC PCB SATELEC...
Страница 228: ...228 SATELEC ON OF SATELEC ON OFF SATELEC PCB ON OFF SATELEC PCB SATELEC SATELEC VIII...
Страница 229: ...229 SATELEC PCB SATELEC XII I SURGE IV I SURGE I SURGE BY DENTAL SATELEC I SURGE I SURGE IEC 60601 1 2...
Страница 232: ...232 XIII 2002 96 23 01 2003 ACTEON XVII XIV BY DENTAL SATELEC XV Ii 93 42 IEC 60601 1 ISO 13485 XVI I SURGE...
Страница 233: ...233 93 42 BF V REF I...
Страница 234: ...234 SN Serial Number...
Страница 237: ...237 Fig 1 Fig 2 10 11...
Страница 238: ...238 Fig 3 Fig 4 Fig 5 Fig 6 Fig 7 Fig 8 Fig 9...
Страница 239: ...239 Fig 10...
Страница 240: ...240...
Страница 241: ...241...