5 Medical device description
5.1 Removing the medical device from its packaging
When you receive your medical device, check for any damage that may have occurred during transportation. If
you have any questions or requirements, contact your supplier.
If you have received this medical device by mistake, please contact the supplier to arrange for it to be
collected.
The AIR-N-GO® easy includes the following items:
l
a handpiece with a non-detachable turbine adapter;
l
a SUPRA 120° nozzle;
l
servicing and maintenance equipment including:
l
a tube of silicone grease,
l
silicone grease applicator brushes,
l
spare O-rings,
l
a syringe and a cannula,
l
a metal shaft cleaning probe,
l
a maintenance instructions leaflet [J10104],
l
a [J10100] Quick Start guide;
l
a [J10101] Quick Clean guide;
l
a starter kit comprising 10 Classic powder sachets and 2 Pearl powder sachets.
Check that the AIR-N-GO® easy adapter is compatible with your quick coupling.
5.2 Installing the medical device
The AIR-N-GO® easy connects directly to the quick coupling on your dental chair.
1. Remove the turbine;
2. Dry the quick coupling using the multi-purpose syringe air function;
3. Do not activate the turbine function when connecting the AIR-N-GO® easy;
4. Remove the lid from the tank;
5. Connect the quick coupling to the AIR-N-GO® easy adapter;
6. Adjust the water flow to drop-by-drop;
7. Wipe the inside walls of the tank with a dry, lint-free cloth;
8. Press the footswitch to eliminate any moisture still in the circuit. Repeat the procedure until there are
no more droplets on the inside walls of the tank;
9. Wipe the inside walls of the tank with a dry, lint-free cloth;
10. Fill the tank with the correct amount of powder for the intended treatment. Fill the tank to the
indicated limit;
11. Wipe with a dry, lint-free cloth to remove all traces of powder from the threads and the tank lid;
12. Check that the O-ring is correctly positioned in the tank lid;
13. Screw the lid back onto the tank;
14. Remove the nozzle and the body from their sterile bags;
15. Install them and start treatment.
If droplets continue to appear on the inside walls of the tank in step 8, read the chapter
Adjust the position of your medical device to correspond to your angle of vision and the characteristics of your
workstation, e.g. lighting or distance between the user and the medical device.
Ensure that your medical device is readily accessible. Do not install your medical device near or on another
device.
Inspect the condition of the O-rings on your quick coupling prior to each use. A damaged O-ring can
cause irreparable damage to your medical device.
Before connecting the AIR-N-GO® easy, dry the chair's quick coupling using the air syringe.
User Manual • AIR-N-GO® easy • J10121 • V5 • 11 • 10/2015 • ND27EN050E - Page
11
/
35
Содержание AIR-N-GO easy
Страница 1: ...User Manual AIR N GO easy...
Страница 6: ......
Страница 8: ...Page 6 35 User Manual AIR N GO easy J10121 V5 11 10 2015 ND27EN050E...
Страница 16: ...Page 14 35 User Manual AIR N GO easy J10121 V5 11 10 2015 ND27EN050E...
Страница 26: ...Page 24 35 User Manual AIR N GO easy J10121 V5 11 10 2015 ND27EN050E...
Страница 28: ...Page 26 35 User Manual AIR N GO easy J10121 V5 11 10 2015 ND27EN050E...
Страница 32: ...Page 30 35 User Manual AIR N GO easy J10121 V5 11 10 2015 ND27EN050E...
Страница 34: ...Page 32 35 User Manual AIR N GO easy J10121 V5 11 10 2015 ND27EN050E...