Model TT339-0700
4-Wire Isolated Frequency Transmitter w/USB
Acromag, Inc. Tel: 248-295-0880
[
30
]
http://www.acromag.com
SPECIFICATIONS
Model Number
Model TT339-0700
•
4 Wire Signal Transmitter
•
Isolated Frequency Input
•
Four-Wire Powered
•
CE Approved
•
Includes UL/cUL Class I,
Division 2 approvals
Custom calibration to
customer specifications can be
added as a separate item at
time of purchase.
Optional factory calibration to customer specifications is ordered as a separate line
item at time of purchase, and on a per unit basis. Factory calibration will require the
specification of input threshold, measurement type, pull-up/pull-down, cutoff
frequency, debounce, samples to average, output update, output range, and I/O
scaling. Normal or reverse acting output and a filter level can also be specified. Use
form 8500-858 for specifying this calibration from our web site at
www.acromag.com.
The standard model without adding custom factory calibration is calibrated by
default to reference test conditions. Recalibration of any transmitter will require use
of a TTC-SIP configuration kit, ordered separately (see Accessories section). Modules
can be mounted on standard 35mm “T” Type DIN rail.
Input
Input Reference Test Conditions:
±5.0V Square Wave; Input Threshold = Bipolar
0.0V; Hysteresis ±28mV; Measurement = Frequency; Pull Up/Down = Disabled;
Cutoff Frequency = 0.5Hz; Samples to Average = 1; Debounce = 0ms; Output Range =
4-
20mA (into 250Ω load); I/O Scaling = 0Hz to 10KHz, Normal Acting; Output Update
= 100ms; Output Update Override = Disabled; Ambient Temperature = 25°C; Power
Supply = 24VDC.
Input Frequency:
Frequency Measurement – 0Hz to 100KHz.
Duty Cycle Measurement – 0Hz to 20KHz.
Input Minimum Pulse Width:
1µs.
Input Acquisition Time:
See Output Update.
Input Pull-up/Pull-down (Internal):
Software selectable 2.7KΩ
input pull-up to +5V
and a 1KΩ
input pull-down to –FRTN, or disabled to disconnect the internal resistors.
CAUTION!
Do not exceed 15VDC at input terminal TB1 with the internal pull-up or
pull-down resistors enabled, or damage to the circuit will occur.
Disable Pull-
up/Pull-Down if transmitter input is AC mains voltage (120V
RMS
)
.
Input Excitation Supply:
+5VDC, current limited to +20mA typical. Bipolar Transient
Voltage Suppressor (TVS) clamps level to ±5.6V typical.
Input Impedance:
37.2