Model TT231-0600
Two-Wire RTD Transmitter w/USB
Acromag, Inc. Tel: 248-295-0880
- 39 - http://www.acromag.com
- 39 -
http://www.acromag.com
USB Interface…
IMPORTANT – USB Isolation is recommended:
The transmitter circuit ground is
connected in common to USB power/signal ground and shield ground. It will make
connection to earth ground when directly connected to the USB port of a Personal
Computer without the use of an isolator. Failure to connect without isolation would
force a potential difference between earth ground at the PC and the earth ground
normally applied in a properly grounded two-wire current loop. This drives an
inadvertent ground loop that can interfere with operation and could damage the
unit. For this reason, USB isolation is strongly recommended when connecting to a
PC. Otherwise, in the absence of USB isolation, a battery powered laptop could be
used to connect to the unit, as the laptop does not normally connect to earth
ground.
Enclosure & Physical
General purpose plastic enclosure for mounting on 35mm “T-type” DIN rail.
Dimensions
: Width = 12.5mm (0.5 inches), Length = 114.5mm (4.51 inches), Depth =
99.0mm (3.90 inches). Refer to Mechanical Dimensions drawing.
I/O Connectors:
Removable plug-in type terminal blocks rated for 12A/250V; AWG
#26-12, stranded or solid copper wire.
Program Connector:
USB Mini B-type, 5-pin. See USB Interface.
Case Material:
Self-extinguishing polyamide, UL94 V-0 rated, color light gray.
General purpose NEMA Type 1 enclosure.
Circuit Board:
Military grade fire-retardant epoxy glass per IPC-4101/98 with humi-
seal conformal coating.
DIN-Rail Mounting:
Unit is normally mounted to 35x15mm, T-type DIN rails. Refer
to the DIN Rail Mounting & Removal section for more details.
Shipping Weight:
0.5 pounds (0.22 Kg) packed.
Environmental
These limits represent the minimum requirements of the applicable standard, but this
product has typically been tested to comply with higher standards in some cases.
Operating Temperature:
-40
C to +80
C (-40
F to +176
F).
Storage Temperature:
-40
C to +85
C (-40
F to +185
F).
Relative Humidity:
5 to 95%, non-condensing.
Isolation:
Input & output are not isolated from each other for this model, as this
model is intended to interface with insulated or non-grounded input sensors.
Installation Category:
Suitable for installation in a Pollution Degree 2 environment
with an Installation Category (Over-voltage Category) II rating per IEC 1010-1 (1990).
Operating Shock & Vibration Immunity:
Sinusoidal Vibration: 5G, 5-500 Hz, in 3 axis
at 2 hours/axis per IEC60068-2-6. Random Vibration: 5G-rms, 5-500 Hz, in 3 axis at 2
hours/axis per IEC60068-2-64. Mechanical Shock: 30g at 11ms half-sine shock pulses
and 50g at 3ms half-sine shock pulses in each direction along 3 axis (18 shocks), per
IEC60068-2-27.