USB Interface 

 

Scans for connected transmitters and allows you to selectively open 
communications with them. 

 

Devices Found:  Connected transmitters are listed in this pull-down 
select menu.  

Click to highlight a transmitter and then click the “Open” 

button to open communications with it. 

 

Open/Close Button:  Click this button to open or close communication 
with the selected transmitter listed in the Devices Found field. 

 

Displays the model number (Product Name), Manufacturer, and Serial 
Number of the connected/selected transmitter. 

 
The software automatically scans for connected transmitters when it is 
booted.  You can select a connected transmitter and open communications 
with it from the Device Found pull-down menu.  The connected/selected 
transmitter’s ID info (Product Name/serial, Manufacturer, & Serial Number) 
is also displayed.  Device connection status messages are also indicated in 
the System Message Window at the bottom of the screen. 

 

Configuration Area 

 

Input Type:  

Set the Input Type, J, K, T, E, R, S, B, N, or ±100mV. 

 

Break Control:  

Set the break detection direction (Up or Down). 

 

Filter Control:  

Set the level of filtering to be applied to the input signal. 

 

CJC Control:  

Select if the CJC is used (On or Off). 

 

I/O Scaling:  

Select the input range endpoints to scale to the 4 to 20mA 

output range endpoints.  You can optionally set this up for a reverse-
acting output by swapping the signals. 

 
 

CONFIGURATION 
SOFTWARE 
 
Introduction 
 

This transmitter can only be 
configured and calibrated via 
its Configuration Software and 
a USB connection to your PC 
or laptop.  The configuration 
software can be downloaded 
free of charge from our web 
site at 

www.acromag.com

 

This software is also included 
on a CDROM bundled with the 
Configuration Kit ST13C-SIP 
(see Accessories section).  
For this model, look for 
program ST132-ST133 
Config.exe. The software is 
compatible with XP or later 
versions of the Windows 
operating system. 
 
The configuration software 
screen for this model is shown 
at left.  The configuration 
screen is divided into five 
sections as follows: USB 
Interface, Configuration, Test, 
Calibration, and the Message 
Bar and controls at the bottom 
of the screen.  Additionally, 
there is a Reset button, a 
button to restore the unit to the 
factory default state, and a 
button to exit the program.  A 
short description of each of 
these groups follows.  For a 
detailed explanation, see 
Configuration Step-by-Step in 
the Technical Reference 
section of this manual. 
 

HELP 

– You can press F1 for 

Help on a selected or 
highlighted field or control.  
You can also click the [?] 
button in the upper-right hand 
corner of the screen and then 
click to point to a field or 
control to get a Help message 
pertaining to the item you 
pointed to. 

Содержание ST132-0600

Страница 1: ...2 0610 Two Wire Transmitter TC Input USER S MANUAL ACROMAG INCORPORATED Tel 248 295 0880 30765 South Wixom Road email sales acromag com Wixom MI 48393 2417 U S A Copyright 2011 Acromag Inc Printed in...

Страница 2: ...libration Connections Reconfiguration Zero Full Scale Calibration Other Configuration Controls SPECIFICATIONS 14 14 16 20 22 24 Model Numbers 24 Input 24 Output 26 USB Interface 27 Enclosure and Physi...

Страница 3: ...il Adapter CE Approved Model ST132 0610 is UL Listed USA Canada suitable for use in Class I Division 2 Groups A B C D Hazardous Locations or Nonhazardous Locations only Model ST132 0610 is ATEX Certif...

Страница 4: ...DIN Form B connection and thermowell protection heads similar to the figure at upper left The M4 mounting screws and relief springs used to attach the transmitter to the connection head are ordered se...

Страница 5: ...hock and vibration applications Since common mode voltages can exist on signal wiring adequate wire insulation should be used and proper wiring practices followed Output wires are normally separated f...

Страница 6: ...n see below Always connect the output power wires and apply loop power before connecting the unit to USB The output of this transmitter has a floating connection relative to ground which makes it flex...

Страница 7: ...ection is typically required and you should connect the metal enclosure s ground terminal green screw to earth ground using suitable wire per applicable codes See the Electrical Connections Drawing fo...

Страница 8: ...model USB ISOLATOR to isolate your USB port or you can optionally use another USB signal isolator that supports USB Full Speed operation 12Mbps Configuration Requires USB and Loop Power This transmit...

Страница 9: ...ol Set the level of filtering to be applied to the input signal CJC Control Select if the CJC is used On or Off I O Scaling Select the input range endpoints to scale to the 4 to 20mA output range endp...

Страница 10: ...s used to calibrate the input output and CJC temperature sensor of the unit Instructions on how to calibrate the unit is detailed in the Configuration Step by Step section of the Technical Reference p...

Страница 11: ...has broken Check your sensor connections to restore operation If connections are OK then your sensor value may be out of range or the unit has been mis calibrated Cannot Communicate with Transmitter v...

Страница 12: ...drop in the wire Ideally the voltage should have ample overhead to drive the load at the maximum output current of 25mA Cannot Calibrate Input Channel Is input wired properly Check input wires at ter...

Страница 13: ...the signals digitally filters the signals and performs analog to digital conversion The A D converter also switches the lead pullups pulldowns to facilitate upscale or downscale thermocouple break det...

Страница 14: ...d junction compensation 2 Wire an output current loop to the transmitter as shown in the Electrical Connections section You will need to measure the output current accurately in order to calibrate the...

Страница 15: ...50 2 585 2 023 2 036 3 048 0 296 0 299 1 340 100 5 269 4 096 4 279 6 319 0 647 0 646 2 774 150 8 010 6 138 6 704 9 789 1 041 1 029 4 302 200 10 779 8 138 9 288 13 421 1 469 1 441 5 913 250 13 555 10...

Страница 16: ...are faded out under these conditions Note that without a device already connected via USB the System Status window at the bottom of the screen prompts you to Connect a ST132 to a USB port After you c...

Страница 17: ...plays the Model the Manufacturer field displays Acromag Inc and the Serial field displays this model s serial number Additionally most of the other fields and controls are not faded out and await your...

Страница 18: ...play a message like No error encountered while reading configuration Select the Input Type T C Type or 100mVDC In the Configuration section of this screen select an input type T C Type J K T R S E B N...

Страница 19: ...the T C input using a voltage source that connects to the module via copper wires as opposed to the wire materials of the various thermocouple types Note CJC temperature values are only resolved to 0...

Страница 20: ...an get a feel for how stable CJC temperatures are by doing this You should turn polling off while trying to calibrate the unit or change its configuration Calibration of Input Output and Tref This sec...

Страница 21: ...0 0 52 410mV Type N TC 0 0 0 000mV 1200 0 43 846mV Type T TC 0 0 0 000mV 390 0 20 255mV Type R TC 0 0 0 000mV 1700 0 20 222mV Type S TC 0 0 0 000mV 1700 0 17 947mV Type E TC 0 0 0 000mV 950 0 72 603mV...

Страница 22: ...its power up conditions e g equivalent to a power on reset This might be useful if you ever encounter erratic operation Return to Out of Box Condition You can use this button to restore the transmitte...

Страница 23: ...configuration No error encountered while sending configuration Getting configuration this takes several seconds Error encountered while reading configuration No error encountered while reading config...

Страница 24: ...enclosures to transmit continuous ambient temperature data via the 4 20mA output Recalibration of any model will require use of an ST13C SIP configuration kit ordered separately see Accessories Models...

Страница 25: ...80 C 0 5 C 1 2 C Input Gain PGA gain is internally adjustable for 1 2 4 8 16 32 64 128 256 and 512 mV mV Note Cold Junction Compensation may be switched off to permit the direct connection of an mV so...

Страница 26: ...C 0 000mV 55661 32768 T 390 C 20 255mV 0 C 0 000mV 50464 32768 R 1700 C 20 222mV 0 C 0 000mV 50432 32768 S 1700 C 17 947mV 0 C 0 000mV 48446 32768 E 950 C 72 603mV 0 C 0 000mV 48626 32768 B 1700 C 12...

Страница 27: ...output span effect per volt DC change USB Interface Includes a USB socket for temporary connection to a PC or laptop for the purpose of setup and reconfiguration During reconfiguration and calibration...

Страница 28: ...The unit can be optionally mounted to 35x15mm T type DIN rails using optional ST130 DIN DIN rail mounting adapter kit Refer to the Mounting Dimensions section for more details Shipping Weight 0 5 pou...

Страница 29: ...ting per IEC 1010 1 1990 Operating Shock Vibration Immunity Sinusoidal Vibration 5G 5 500 Hz in 3 axis at 2 hours axis per IEC60068 2 6 Random Vibration 5G rms 5 500 Hz in 3 axis at 2 hours axis per I...

Страница 30: ...s manual for detailed information on available software control of this model Software Interface Package Configuration Kit Order ST130 SIP USB Signal Isolator USB A B Cable 4001 112 USB A mini B Cable...

Страница 31: ...your PC and the USB isolator It is normally included with the ST13C SIP Software Interface Package and also with the isolator model USB ISOLATOR USB A mini B Cable Order 4001 113 USB A mini B Cable 4...

Страница 32: ...g com https www acromag com 32 Revision History The following table shows the revision history for this document Release Date Version EGR DOC Description of Revision 03 DEC 13 J FJM ARP Change the Def...

Отзывы: