MiniPuck Model ST132-06x0 Two-Wire TC/mV Transmitter
_________________________________________________________________________________
______________________________________________________________________________________
Acromag, Inc. Tel: 248-295-0880 Email:
22
Calibration
– Restore Factory Calibration
If you make an error in recalibration, such that the unit has degraded
performance, then you can click the “Restore Factory Calibration” button to
restore the unit to the original factory calibration. Note that doing so will not
change/restore any configuration settings, only the range calibrations. You
sh
ould only click the “Restore Factory Calibration” button if you ever mis-
calibrate or misconfigure a transmitter in such a way that its operation
appears erratic.
Reset ST13x Button
You can use
“Reset ST13x”
to reset the transmitter and cause it to revert to
its power-up conditions (e.g. equivalent to a power-on reset). This might be
useful if you ever encounter erratic operation.
Return to Out-of-Box Condition
You can use this button to restore the transmitter configuration to the
original factory state (see Specifications Reference Test Conditions). This
control provides a potential recovery path should the configuration ever
become corrupted during recalibration, perhaps due to miscalibration. For
example, if during calibration you break the USB connection before
completing calibration, a memory transfer checksum value could be
corrupted, and this would inhibit normal operation. Alternately, this button
can be used as a sanitation tool to restore the unit to its initial configuration.
Note that the “Restore Factory Calibration” control of Calibration just affects
the calibration of the unit, different from this control which sends the unit to
its initial factory configuration and calibration. This button will restore
ALL
configuration settings and calibration to their original factory settings.
CONFIGURATION
STEP-BY-STEP
Other Configuration
Controls