Model SP33X-0700
4-Wire DC I/V Signal Splitter w/USB
Quick Overview
–
Windows...
CALIBRATION...continued
Input…
Before attempting calibration, first set the Input Range
to calibrate from the I/O Config/Test page and be sure to
click the
[Send I/O Config]
button. On the Calibration
page, click the
[Input Cal Instructions]
button to begin
input calibration.
When you click the
[Zero]
or
[Full Scale]
buttons of the
Input Calibration section, you will be prompted to apply
a specific current level at TB1, or voltage level at TB2,
depending on your selected input range. Once you have
applied this signal to the correct input terminals, click
the
[OK]
button of the prompt and follow the on-screen
instructions to complete input calibration.
Output…
Click the
[Output Cal Instructions]
button to begin
output calibration. You will be prompted to adjust the
input signal as required to drive the output to its precise
output range zero or full-scale level. Then once the
output is set to zero or full-scale, you simply click the
corresponding
[Zero]
or
[Full-Scale]
button of the
CALIBRATION - Output section to set the output range
zero or full-scale endpoint.
Factory Settings
(Use in Case of Trouble or for Sanitation Purposes)
•
Restore a transmitter to its original factory
calibration.
•
Restore a transmitter to its initial factory
configuration.
You can click the “Restore Factory” buttons
if you ever
misconfigure or improperly calibrate a transmitter such
that its operation appears erratic, or for sanitation
purposes when decommissioning a module.
Calibration Status (Bottom of Screen)
•
Displays communication status messages for the
calibration process.
The CALIBRATION STATUS message bar at the bottom of
the screen will display status messages relative to
calibration.