Quick Overview
–
Android Reconfiguration...continued
Input/Output Configuration
The I/O Configuration screen is shown at left and is used to Configure your
splitter Input, splitter Outputs, and AUX output. Scroll down this screen to
set your input type/range, input digital filtering level, rescale each output,
enable and scale the AUX output, and even scale the input differently for
each output via this screen.
Note that if your unit is connected when you select this tab, the app
aut
omatically reads your splitter’s current I/O and scaling information and
displays it.
Likewise, changing any option on this page sends the changes to the
splitter immediately.
Note that the Device Status is indicated at the bottom of all pages and will
report if changes were sent successfully (Connected).
Input Calibration
If you have correctly setup your unit and encounter excessive error, you
may click the Calibration tab to display the Calibration control screen
shown at left, which presents Input calibration controls first, followed by
Output calibration controls as you scroll down the page.
IMPORTANT:
The splitter has had its input & output channels factory
calibrated with high precision. Attempting to recalibrate the input or
outputs may degrade performance if done improperly, or using low grade
equipment. Consider recalibration carefully.
The selected input range is indicated at the top and this is the range that
will be calibrated. The software does not use your scaled range zero to
calibrate, but the zero of the nominal input range selected. Some sub-
ranges have their calibration extrapolated from the calibration of a larger
native range. Always calibrate the largest native range first to keep its
recalibration from over-writing any sub-range calibration.
For input zero calibration, connect a precise input signal level for the zero
of your range, then tap the
[CALIBRATE INPUT ZERO]
button one time to
set the input ADC level to its input range zero (0%) point.
For full-scale calibration, connect a precise input signal level for the full-
scale value of your range, then tap
[CALIBRATE INPUT FULL-SCALE]
one
time to set the input ADC level to its input range full-scale (100%).
The device status at the bottom of the page will report if the calibration
was sent successfully.