Model DT23x-0600
Dual Two-Wire DC I/V Input Transmitters
Acromag, Inc. Tel: 248-295-0880
- 19 -
http://www.acromag.com
- 19 -
https://www.acromag.com
Configuration
Communication Connection (Start Here)
After executing the Acromag Configuration software for your
specific model, a Communication Setup screen like that shown
at left will appear
if you have not already connected to your
transmitter via USB
(note fields are blank and a red status
message appears).
Connect your Windows PC to the unit via isolated USB and its
model-serial information will appear in the Device select scroll
field as shown in the second screen at left.
Note that if you have a connection path to more than one unit
via a USB hub, you can use the Device scroll field to select a
different unit using the serial number suffix of the Device Model
to discern one unit from another.
Once you have selected a device/serial, click the
[Open]
button
to open communication with the unit.
After clicking
[Open]
, the selected unit’s Model, Serial Number,
Manufacturer, and connection Status will be displayed as
shown in the third screen at left.
After you conne
ct USB and “Open” communicati
on with a unit,
the Status field indicates “Device opened successfully”
as
shown in third screen at left.
At this point, you can click the “
Configuration
” tab to begin
configuring the unit,
“Calibration” to calibrate an input or
output channel, or “Diagnostics”
to optionally test its operation
by polling its input channels.
Note that you must have power connected to output loop 1 of
the transmitter, even if connected via USB. If you are in dual
transmitter or splitter mode, you must have power connected
in both output loops of the unit. You will not be able to
configure, calibrate, or test a unit without loop power also
applied to at least output 1.
Note that you must have power connected to output loop 1
of the transmitter in any operating mode, even if
connected to USB. If you are in dual transmitter or splitter
mode, you must have power connected in both output
loops of the unit. You will not be able to configure,
calibrate, or test a unit without loop power also applied to
at least output 1.