Serinus 60 User Manual 3.0
Page 26
Mixer Preamp PCA
Amplifies the optical signal received by the vacuum photodiode detector. The resulting signal is
converted to a frequency and transmitted to the Interface PCA.
1.5.9
Pneumatic Tubing
The pneumatic tubing inside this instrument is specially designed for use in Ecotech Serinus
instruments. It has the flexibility of Tygon tubing with the added inner sheath of Teflon to prevent
contamination of the sample. Care should be taken when removing and inserting the tubing into the
barbed fittings.
Figure 3 – Ecotech Tygon Tubing
1.5.10
Main Controller PCA
The main controller PCA controls all the processes within the instrument. As well as the on-board
microprocessor, it contains a battery backed clock, calendar, analog to digital converters and many
other circuits for signal processing and control. The ambient pressure and chassis temperature
sensors are also located on this board. The main controller PCA is located above all other
components within the instrument. It pivots on hinges to allow access to the components
underneath.
CAUTION
Never place objects on top of the main controller PCA as it may result in damage.
1.5.11
Power Supply
The power supply is a self-contained unit housed in a steel case designed to meet all the relevant
safety and EMC requirements. This new revision of power supply is different to the previous revision
as there is no need to set the operating voltage switch because it is autoranging.
The output of the power supply pr12V, +5V, -12V and +3.3V to the instrument.
1.5.12
On/Off Switch
The on/off switch is located on the back panel (bottom right facing the rear of the instrument). It is
part of the power supply.
Содержание Met One Instruments GAS-1060
Страница 10: ...Serinus 60 User Manual 3 0 Page 10 This page is intentionally blank...
Страница 18: ...Serinus 60 User Manual 3 0 Page 18 This page is intentionally blank...
Страница 35: ...Installation Page 35 Figure 7 Switching the Battery On Off...
Страница 36: ...Serinus 60 User Manual 3 0 Page 36 This page is intentionally blank...
Страница 102: ...Serinus 60 User Manual 3 0 Page 102 This page is intentionally blank...
Страница 122: ...Serinus 60 User Manual 3 0 Page 122 7 6 USB Memory Stick Failure Figure 64 USB Memory Stick Failure...
Страница 134: ...Serinus 60 User Manual 3 0 Page 134 9 5 Plumbing Schematic PN D020060...
Страница 135: ...Parts List Page 135 9 6 Block Wiring Schematic PN D020111...
Страница 136: ...Serinus 60 User Manual 3 0 Page 136 9 7 Measurement Cell Side Section PN H015000...
Страница 137: ...Parts List Page 137...
Страница 138: ...Serinus 60 User Manual 3 0 Page 138 9 8 Calibration Valve Manifold Assembly Exploded View PN H010013 01...
Страница 139: ...Parts List Page 139 9 9 Valve Assembly Exploded View PN H010042...
Страница 140: ...Serinus 60 User Manual 3 0 Page 140 9 10 High Pressure Valve Exploded View PN H050043...