Introduction
Page 19
1.
Introduction
1.1
Description
The Serinus 60 Nitrogen Dioxide Analyser (NO2) utilizes cavity attenuated phase shift spectroscopy
detection to perform continuous analysis of nitrogen dioxide.
The Serinus 60 can measure to a sensitivity of 20 ppt with a range of 0-1 ppm. The measurement of
this gas is performed with the following main sub-components:
Particulate filters
Sample dryer
Measurement cell (temperature controlled)
Light source (Blue LED) and Photo Detector
Interface PCA
A microprocessor programmed with Serinus firmware monitors the detector response and many
other parameters. The concentrations are automatically corrected for gas temperature and pressure
changes and when using gravimetric units (e.g. μg/m³) is referenced to 0°C, 20°C or 25°C at 1
atmosphere. This allows the Serinus 60 to provide readings in the relevant units for your
requirements.
This section will describe the specifications of the instrument as well as the main
components and techniques used to obtain stable gas concentration readings.
1.2
Specifications
1.2.1
Measurement
Range
0-1 ppm autoranging
Lower detectable limit: 0.04 ppb, with no filter active
Noise:
< 0.02 ppb
1.2.2
Precision/Accuracy
Precision
0.04 ppb otherwise 0.5 % of reading, whichever is greater
Linearity
±4 % of full scale
Response Time
30 seconds
Содержание Met One Instruments GAS-1060
Страница 10: ...Serinus 60 User Manual 3 0 Page 10 This page is intentionally blank...
Страница 18: ...Serinus 60 User Manual 3 0 Page 18 This page is intentionally blank...
Страница 35: ...Installation Page 35 Figure 7 Switching the Battery On Off...
Страница 36: ...Serinus 60 User Manual 3 0 Page 36 This page is intentionally blank...
Страница 102: ...Serinus 60 User Manual 3 0 Page 102 This page is intentionally blank...
Страница 122: ...Serinus 60 User Manual 3 0 Page 122 7 6 USB Memory Stick Failure Figure 64 USB Memory Stick Failure...
Страница 134: ...Serinus 60 User Manual 3 0 Page 134 9 5 Plumbing Schematic PN D020060...
Страница 135: ...Parts List Page 135 9 6 Block Wiring Schematic PN D020111...
Страница 136: ...Serinus 60 User Manual 3 0 Page 136 9 7 Measurement Cell Side Section PN H015000...
Страница 137: ...Parts List Page 137...
Страница 138: ...Serinus 60 User Manual 3 0 Page 138 9 8 Calibration Valve Manifold Assembly Exploded View PN H010013 01...
Страница 139: ...Parts List Page 139 9 9 Valve Assembly Exploded View PN H010042...
Страница 140: ...Serinus 60 User Manual 3 0 Page 140 9 10 High Pressure Valve Exploded View PN H050043...