Stationary Photoionization Detector SPID2
User Manual
SPID2 - User Manual (en) - 1.11.docx
Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de
Version 1.11
Page 45 of 60
7.7.
How To
–
Replace the Filter Port
–
Filter Cartridge
The filter cartridge must be replaced, if the error
message “Filter dirty” wil
l be shown on display.
Remove Filter Port
–
Filter Cartridge
1.
Unscrew the old filter cartridge counter-
clockwise.
Replace Filter Port
–
Filter Cartridge
2.
Check that the filter port
–
seal is fitted and
not damaged.
3.
Check that the filter cartridge
–
seal is not
broken.
4.
Screw the filter cartridge into the filter port
unit.
Only light force is required!!!
5.
Open the filter cartridge after installation by
using the opener.
For detailed information, see chapter 4.3.
Calibrate the device after installation of a new filter cartridge.