ACE INSTRUMENTS ADVISES
EVEN IF YOU GET A RESULT MEANING „SAFE RANGE“, YOU SHOULD NEVER DRIVE AFTER
CONSUMING ALCOHOL OR DO ANY OTHER ACTIVITIES THAT ARE SEVERELY IMPAIRED BY
ALCOHOL CONSUMPTION AND MIGHT ENDANGER YOUR HEALTH AND SAFETY OR THE
HEALTH AND SAFETY OF OTHERS.
Status Indicators on the Display
#82 D365
Total number of all tests performed
(will be reset to 0 after a calibration)
Remaining days until
calibration
Remaining battery
power
Display Indications
NOTICE
MEANING
ACTION
BLOW ERROR
The sampling volume or blow
power is not enough to analyze
the breath sample.
Press the power button to perform
a new test.
LOW BATTERY
The testing cannot be
performed because the
battery is low.
Replace the integrated batteries
with new batteries.
CALIBRATION DUE
DATE HAS PASSED
The recommended calibration
is due. This display may
depend on the time factor
(default setting: 12 months). The
setting can be changed under
“Advanced Menu Settings”.
Contact your local distributor or
customer service center for proper
recalibration service.
OVER NUMBER
OF TEST
ALLOWED
The recommended calibration
is due. This display may
depend on the number of tests
(default setting: 500 tests). The
setting can be changed under
“Advanced Menu Settings”.
Contact your local distributor or
customer service center for proper
recalibration service.
BREATH SAMPLE
WITH EXCESSIVE
SALIVA
Excessive saliva went into the
sensor during the test.
Wait 20 minutes, and then try again.
If this error code is displayed for a
long time, contact the ACE Instru-
ments Service Center
(see www.alkomat.net)
OUT OF TEMP
RANGE
The testing cannot be per-
formed because the device
is outside of recommended
operating temperature.
Repeat the test after the device has
reached a temperature within the
operating range (-5°C up to 40°C).
NO MEMORY
CAPACITY
AVAILABLE
No further test can be per-
formed because the number
of tests has already reached
10,000 and no calibration has
been performed so far
(10,000 = maximum number of
tests possible without a single
calibration).
Contact your local distributor or the
ACE Instruments Service Center
(see www.alkomat.net).
Basic Menu Options
How to enter the menu options: During the warm-up process, press the menu button for
2 seconds. You will see these menu options:
TEST RESULT
TEST RESULT-G
ALARM LIMIT
UNIT
BRIGHTNESS
SOUND
INFORMATION
LANGUAGE
PIN CODE
How to move within the menu options:
Š
To move the cursor up, quickly press the power button.
Š
To move the cursor down, quickly press the menu button.
Š
To enter the selected option, press the power button for 2 seconds.
Š
To save the selected setting, press and hold the power button and menu button
simultaneously for 2 seconds
Š
To return to warm up process, press the menu button for 2 seconds.