69
Setup
Options
Probe Control
The Probe Control options allow you to change the
default settings of the probe. The options are:
Frequency
Depth
Cine Buffer
Start/Stop
Power
Frequency
The probe frequency field allows you to select from two
frequencies: 12 MHz or 15 MHz. Which frequency you
choose for scanning will determine what depth settings
you can select. If you choose 12 MHz the depth options
are 30 mm and 60 mm. If you choose a frequency of 15
MHz the depth options are 50 mm and 100 mm. The
default probe settings are 12 MHz and 60 mm.
To change the frequency select the appropriate value
from the drop down list located at the right of the
frequency field as shown in Figure 51 below.
Figure 51
Selecting Probe Frequency
Содержание B-Scan Plus
Страница 1: ...Accutome B Scan Plus User s Guide 24 6110 Rev D...
Страница 20: ...xviii Accutome B Scan Plus User Guide...
Страница 26: ...6 Accutome B Scan Plus User Guide...
Страница 54: ...34 Accutome B Scan Plus User Guide Figure 16 TGC Tab Active Figure 17 Gamma Tab Active...
Страница 108: ...88 Accutome B Scan Plus User Guide...
Страница 114: ...94 Accutome B Scan Plus User Guide Figure 74 Initial Exam Screen with Patient Name...
Страница 173: ...153 Patient Image Analysis TGC Adjustment Figure 118 TGC Adjustments Displayed...
Страница 192: ...172 Accutome B Scan Plus User Guide...
Страница 196: ...176 Accutome B Scan Plus User Guide Figure 136 Printing a Frame PRINT IMAGE...
Страница 209: ...189 Printing Images and Reports Creating Reports...
Страница 210: ...190 Accutome B Scan Plus User Guide...
Страница 254: ...234 Accutome B Scan Plus User Guide...
Страница 256: ...236 Accutome B Scan Plus User Guide...
Страница 268: ...248 Accutome B Scan Plus User Guide...