The ProPanel Readout Programming Parameters are listed below.
Values in
[ ]
are the range of values available for that Parameter.
Factory Default Value is shown in
Bold
Red.
Parameters that are CUSTOM SET for ProPanel operation are:
Parameter Pr2
has been programmed to
1
Parameter Pr11
has been programmed to
2
Parameter Pr13
has been programmed to
Custom value per System
Parameter Pr14
has been programmed to
0
Parameter Pr30
has been programmed to
1
Parameter Pr31
has been programmed to
1.000in
Parameter Pr32
has been programmed to
1.500in
Parameter Pr31
has been programmed to
0.750in
Parameters that
DO NOT
apply to ProPanel systems are shown
Grayed Out.
NOTE: Programming parameters are not sequentially numbered due to
firmware differences between systems and provision for future
enhancements / features.
Pr1
– Datum Key
[0 to ± 999.999in] or [0 to ± 9999.99mm]
The value that will be recalled (on the readout) when the
DATUM
key is
pressed. This value would normally be set to 0.00, (Jaws closed) but may be
set to any value the readout is capable of displaying.
Default Value= 0.00
Pr 2
– Direction of Travel
[0 or 1]
This parameter controls the sign of travel (positive vs. negative) when the
measuring system is moved.
FACTORY SET FOR PROPANEL TO 1
Pr3
– Key Lockout
[0 or 1]
This parameter controls the operation of the
+
,
-
and
DATUM
keys.
Default Value = 0
Pr 3 Setting
Key
Action
0
+, - and Datum keys respond normally
1
+, - and Datum keys are locked