AM8040 Precision Thermometer User’s Guide
Rev. 202206V1
14/40
5.2
25 Ω calibration
5.2.1
The process of 25 Ω calibrations is similar to that of 100 Ω
calibrations. First, turn off the power of the instrument and
ensure the measurement/calibration switch at the rear panel is set
to the “CAL” side (right side), as shown in Figure 4. Turn on the
power.
5.2.2
Choose 25 Ohm CALIBRATION under the main menu
(*CALIBRATION MENU*) by pressing the
or
keys to
move the “>” cursor to the “25 Ohm CALIBRATION” line:
* CALIBRATION MENU
100 Ohm CALIBRATION
> 25 Ohm CALIBRATION
Enter = Yes
-------------------------------------------
5.2.3
Press the
key to start the program of 25 Ohm
CALIBRATION. The instrument will show:
* 25 Ohm CALIBRATION
Connect a 25 Ohm
Standard Resistor
and Press Enter
Menu= Exit Enter = Yes
-------------------------------------------