AM1760/1762 Secondary Standard 

Platinum Resistance Thermometer 

User’s Guide

 

 
 
 
 
 

 

 
 
 
 
 
 

 

 

Please download the latest version at www.accumac.com 

 

Содержание AM1760

Страница 1: ...Rev 201609V3 AM1760 1762 Secondary Standard Platinum Resistance Thermometer User s Guide Please download the latest version at www accumac com...

Страница 2: ...brations 3 2 Specifications 4 2 1 Specifications 4 3 General Operations 5 3 1 Connecting to the readout device 5 3 2 Drive Current 5 3 3 Stability of Readings 5 3 4 Immersion Requirements 5 3 5 Therma...

Страница 3: ...ve component 5 Follow all other safety guidelines listed in this user s guide Cautions Follow these guidelines to avoid possible damage to the instrument 1 Avoid mechanical shocks DO NOT drop or slam...

Страница 4: ...designed as a secondary standard thermometer to calibrate other thermometers and to measure temperature precisely in metrology labs 1 2 Main Features High accuracy 0 006 C at 0 01 C Temperature range...

Страница 5: ...o step change in water moving at 3 feet per second Measurement Current 1 mA Sensor Length 42 mm Sensor Location 5 mm from tip Insulation Resistance 1000 M at room temperature Sheath Material Inconeltm...

Страница 6: ...ach end of the sensor are identified by red black and white blue insulation 3 2 Drive Current AccuMac recommends 1mA as drive current to ensure the best measurement 3 3 Stability of Readings To achiev...

Страница 7: ...gone through an annealing process and stability test to minimize the thermal EMF which is caused by either impurities of sensing element or temperature differentials at lead wires connection point 3 6...

Страница 8: ...ible b When shipping the PRT use protective box and other protective packaging materials to minimize mechanical shocks as much as possible c When using dry blocks make sure the well diameter is approp...

Страница 9: ...rvice Be sure to have the model number and serial number of your PRT available 5 2 Problem Causes and Solutions a R0 Rtpw becomes higher significantly This is likely caused by mechanical shocks The PR...

Страница 10: ...ty terms in the separate agreement shall control Each product from AccuMac Corporation is warranted to be free from defects in material and workmanship under normal use and service The warranty period...