Set up Menu / Configuring Scale

(Factory set parameters are marked “0”.)
Turn off scale.
Turn on while holding the ZERO/TARE key.
Page or level value “1” will be displayed.
To navigate within a menu level press “F”, when you reach the last menu item, the first   item is shown again.
To select the next line or level within a group press and release “CAL/PRINT” key.
To return to the previous level (or page) press “ZERO/TARE” Key.
To confirm selected parameters press and hold the “CAL/PRINT” key for more than 3 seconds.
“0” indicates the selected menu setting.
To store selected codes and exit menu press and hold “ZERO/TARE” key for 3 seconds or until display 
returns to weigh mode.

6

8

7

9

Unit Toggling

Auto Zero

Calibration Setting

Baud Rate

Parity

141 Unlocked “0”
142 Locked

151 On “0”
152 Off

171 Calibration on “0”
172 Calibration off

211 1200 Baud “0”
212 2400 Baud
213 4800 Baud
214 9600 Baud

221 Mark
222 Space
223 Odd “0”
224 Even

No. Of Stop Bits

Manual / Auto Print Mode

Disable Keys 

(Disables All Keys Except for “ZERO/T

TARE” Key)

Battery ON/OFF Mode

Factory Settings

231 1 “0”
232 2

311 Manual without stability
312 Manual after stability
313 Auto without stability
314 Auto at stability

411 On
412 Off “0”

511 Turns off after 4 minutes of in-activity “0”
512  Always on
513 Turns off after 10 minutes of in-activity

911 Reset

Scale Operating Menu:

Содержание SVI-100E

Страница 1: ...PROFESSIONAL PRECISION INDUSTRIAL SCALE SVI SERIES SVI 50C SVI 100E SVI 200F SVI 10A SVI 20B...

Страница 2: ...ion Programs Features Care Maintenance Menu Settings Scale Operating Manu Interface Description Product Specifications AC Adapter Battery Operation On Off Cal Print F Key Zero Tare Light Weight Unit C...

Страница 3: ...recommended to remove battery when scale is not in use for extended periods Also replace low or dead batteries Note Back lit display is off unless activated with Light key ACAdapter Battery Operation...

Страница 4: ...cription Calibration Weight Calibration with the pan cleared and zero stability indicated in display press and hold the CAL PRINT key for more than 3 seconds to enter the calibration mode Depending up...

Страница 5: ...t options g lbs ozt kg depending upon model Select desired weigh unit by pressing CAL PRINT selected option will be displayed and balance will return to weighing mode Pressing the F key will enable yo...

Страница 6: ...th water or cleaning liquids Please use supplied packaging when transporting scale Remember to remove battery when not in use 1 2 3 5 4 Totaling To select the application Totaling press and hold TARE...

Страница 7: ...ore selected codes and exit menu press and hold ZERO TARE key for 3 seconds or until display returns to weigh mode 6 8 7 9 Unit Toggling Auto Zero Calibration Setting Baud Rate Parity 141 Unlocked 0 1...

Страница 8: ...Totaling Mode Calibration Mode ASCII p t u f a b c D HEX 70 0D 0A 74 0D 0A 75 0D 0A 66 0D 0A 61 0D 0A 62 0D 0A 63 0D 0A 64 0D 0A CR LF CR LF CR LF CR LF CR LF CR LF CR LF CR LF Output Format Position...

Страница 9: ...ozt 100 kg 220 lbs 3215 ozt 200 kg 440 lbs 6430 ozt 1 g 0 001 kg 0 002 lbs 0 05 ozt 2 g 0 002 kg 0 005 lbs 0 1 ozt 0 005 kg 0 01 lbs 0 2 ozt 0 02 kg 0 05 lbs 0 5 ozt 0 05 kg 0 1 lbs 1 ozt grams kilog...

Страница 10: ...Zone 100102 Bejing PR China TEL 86 10 6439 2233 If you have any questions please contact your local Acculab supplier visit our website www acculab com or contact Si tiene alguna duda p ngase en contac...

Отзывы: